| Literature DB >> 1629345 |
W Dumornay1, P M Roblin, M Gelling, M R Hammerschlag, M Worku.
Abstract
The performance of Magic Lite (CIBA-Corning), a chemiluminescent immunoassay (CIA), was compared with that of culture for the diagnosis of neonatal chlamydial conjunctivitis and respiratory infection. We performed a retrospective evaluation of 51 ocular and 96 nasopharyngeal specimens previously collected for culture testing with the CIA. The sensitivities for the ocular and the nasopharyngeal specimens were 91 and 91.7%, respectively. The specificities for both sites were 100%. A subsequent prospective study evaluating 71 ocular and 38 nasopharyngeal specimens revealed sensitivities of 83.3 and 20%, respectively. The specificities for both sites were 100%. The CIA performed favorably, compared with culture, for the diagnosis of chlamydial conjunctivitis; however, the CIA appeared less sensitive for the diagnosis of respiratory infection, including pneumonia.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1629345 PMCID: PMC265395 DOI: 10.1128/jcm.30.7.1867-1869.1992
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948