Jens Hagemeister1, Frank M Baer, Robert H G Schwinger, Hans W Höpp.
Abstract
BACKGROUND: Cobalt chromium coronary stents are increasingly being used in percutaneous coronary interventions. There are, however, no reliable data about the characteristics of unfolding and visibility of this stent alloy in vivo. The aim of this study is to compare cobalt chromium coronary stents with conventional stainless steel stents using intracoronary ultrasound.Entities:
Year: 2005 PMID: 16255780 PMCID: PMC1283148 DOI: 10.1186/1468-6708-6-17
Source DB: PubMed Journal: Curr Control Trials Cardiovasc Med ISSN: 1468-6694
Baseline Lesion Characteristics (QCA).
| Preprocedure reference vessel diameter (mm) | 3.04 ± 0.34 | 2.99 ± 0.47 |
| Preprocedure MLD (mm) | 0.77 ± 0.32 | 0.63 ± 0.32 |
| Preprocedure diameter stenosis (%) | 73.50 ± 10.30 | 79.60 ± 8.80 |
| Target vessel | ||
| Left anterior descending | 20% | 50% |
| Circumflex | 20% | 20% |
| Right coronary artery | 60% | 30% |
| Postprocedure MLD (mm) | 3.34 ± 0.34 | 3.31 ± 0.48 |
| Postprocedure diameter stenosis (%) | 6.00 ± 1.80 | 5.10 ± 3.80 |
Figure 1Ballon pressure by IVUS-criteria: balloon deployment pressure when IVUS criteria were reached for Multi-Link Vision® and Multi-Link Zeta® (n = number of lesions).
Visibility-Index: generated from QCA- and IVUS-data.
| Postprocedure MLD QCA (mm) | 3.32 ± 0.41 | 3.34 ± 0.34 | 3.31 ± 0.48 |
| Postprocedure MLD IVUS (mm) | 3.07 ± 0.41 | 2.95 ± 0.17 | 3.19 ± 0.55 |
| Visibility [QCA/IVUS] | 1.08 | 1.13 | 1.04 |