Literature DB >> 16251566

Esophageal Doppler and thermodilution are not interchangeable for determination of cardiac output.

Sophie Collins1, François Girard, Daniel Boudreault, Philippe Chouinard, Louis Normandin, Pierre Couture, Marie-Josée Caron, Monique Ruel.   

Abstract

PURPOSE: This study compares thermodilution cardiac output (TD-CO) and esophageal Doppler cardiac output (ED-CO) during periods of hemodynamic stability and after heart stabilization during off-pump coronary artery bypass (OPCAB) surgery.
METHODS: After Institutional Review Board approval, 58 patients undergoing OPCAB had simultaneous comparison of TD-CO and ED-CO at three time periods. Measurements were recorded, in a blinded manner, after probe insertion (T0), immediately before and after (T1,T2) heart displacement and before starting any pharmacological treatment (if needed) to maintain systolic blood pressure to its value before heart mobilization. Measurements were also taken before sternal closure (Tfinal).
RESULTS: Three hundred and two pairs of data were analyzed using the Bland and Altman method. Bias, standard deviation (SD) of the bias (precision), and degree of agreement (bias +/- 2 SD) were calculated. Based on published literature, we considered that the highest degree of agreement should be < 0.5 L.min(-1) to consider both methods as interchangeable. At T0, bias and SD of bias between TD-CO and ED-CO were -0.1 +/- 1.0 L.min(-1). Immediately before heart stabilization, bias +/- SD was 0.6 +/- 1.0 L.min(-1) and after heart displacement, 0.5 +/- 0.8 L.min(-1). At Tfinal, bias +/- SD was 0.7+/- 0.7 L.min(-1).
CONCLUSION: Because the degree of agreement was > 0.5 L.min(-1) at all measurement periods except T0, we conclude that TD and ED are not interchangeable at any time during OPCAB surgery.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16251566     DOI: 10.1007/BF03022062

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  3 in total

Review 1.  Transesophageal Doppler devices: A technical review.

Authors:  Patrick Schober; Stephan A Loer; Lothar A Schwarte
Journal:  J Clin Monit Comput       Date:  2009-10-20       Impact factor: 2.502

Review 2.  Minimally invasive monitoring of cardiac output in the cardiac surgery intensive care unit.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Giorgio della Rocca; Maxime Cannesson; Christoph K Hofer
Journal:  Curr Heart Fail Rep       Date:  2010-09

3.  Invasive and noninvasive cardiovascular monitoring options for cardiac surgery.

Authors:  Dominic P Recco; Nathalie Roy; Alexander J Gregory; Kevin W Lobdell
Journal:  JTCVS Open       Date:  2022-04-11
  3 in total