Literature DB >> 1620887

Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch.

W De Neve1, F Van den Heuvel, M De Beukeleer, M Coghe, L Thon, P De Roover, M Van Lancker, G Storme.   

Abstract

We have evaluated the fluoroscopic on-line portal imaging (OPI) system developed by Siemens (Beamview-1, Concord, CA, U.S.A.) in routine clinical radiotherapy, involving the treatment of 883 fields (559 patient set-ups for treatment) on 21 patients. The image was typically generated by delivering 10 monitor units when used in single exposure or 1-2 monitor units on a large open field followed by 8-10 monitor units on the actual field when double exposure was used. Comparison between the portal image and the simulator film was done by eye. A region of tolerance was drawn on the simulator film and the field edges on the portal image had to project within this region. If this criterion was not met, adjustments followed by verification portal images were done before the remaining field dose was delivered. If possible, these adjustments were performed by moving the patient couch by remote control. The image quality was insufficient for evaluation in 75/883 (8.5%) fields. The abovementioned criterion was not met in 95/808 (11.8%) of the evaluable fields (26/559 patient set-ups were not evaluable). Of the 533 evaluable patient set-ups, 92 had to be adjusted (17.2%) including three (pelvic irradiations) set-ups that were adjusted on both field irradiated during the same radiotherapy session. In one case an incorrect tray (with wrong blocks) was detected and replaced. In one case (a 5.5 x 6.0 cm rectangular larynx field) the x and y axis of the field were interswitched. In one case incorrect focusing of a block was shown by the portal image. To make adjustments, the couch longitudinal position was changed 20 times (range -10 to +15 mm). The lateral position was changed 73 times (range -15 to +16 mm). The height position was changes 6 times (range -7 to +6 mm). Diaphragma rotation changes were performed 5 times (1 degree). The fraction of treatment time that was related to the use of OPI was 30.7% median (mean 32.4%, S.D. 14.1%). The range was 4.1 to 78.6%. On the basis of calculations assuming no OPI would have been used, field treatment time was increased by a median of 44.2% (mean 55.8%; S.D. 41.2%) by using OPI. The fraction of monitor units (fraction of the dose) to generate a satisfactory image was 10% median.(ABSTRACT TRUNCATED AT 400 WORDS)

Entities:  

Mesh:

Year:  1992        PMID: 1620887     DOI: 10.1016/0167-8140(92)90353-v

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  7 in total

1.  Uncertainties and CTV to PTV margins quantitative assessment using cone-beam CT technique in clinical application for prostate, and head and neck irradiation tumours.

Authors:  X J Juan-Senabre; J López-Tarjuelo; A Conde-Moreno; A Santos-Serra; A L Sánchez-Iglesias; J D Quirós-Higueras; N de Marco Blancas; S Calzada-Feliu; C Ferrer-Albiach
Journal:  Clin Transl Oncol       Date:  2011-11       Impact factor: 3.405

Review 2.  The evolving role of stereotactic radiosurgery and stereotactic radiation therapy for patients with spine tumors.

Authors:  Jack P Rock; Samuel Ryu; Fang-Fang Yin; Faye Schreiber; Muwaffak Abdulhak
Journal:  J Neurooncol       Date:  2004 Aug-Sep       Impact factor: 4.130

3.  CTV to PTV margins for prostate irradiation. Three-dimensional quantitative assessment of interfraction uncertainties using portal imaging and serial CT scans.

Authors:  Luis A Pérez-Romasanta; Eva Lozano-Martín; Joaquín Velasco-Jiménez; Fermín Mendicote-León; Miguel Sanz-Martín; Javier Torres-Donaire; Carmen Carrascosa-Fernández; Juan Carlos Zapata-Jimínez; Jacinto Arjona-Gutiérrez; Antonio Gil-Agudo
Journal:  Clin Transl Oncol       Date:  2009-09       Impact factor: 3.405

4.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

5.  Impact of body-mass factors on setup displacement in patients with head and neck cancer treated with radiotherapy using daily on-line image guidance.

Authors:  Yo-Liang Lai; Shih-Neng Yang; Ji-An Liang; Yao-Ching Wang; Chun-Yen Yu; Ching-Hsiung Su; Shang-Wen Chen
Journal:  Radiat Oncol       Date:  2014-01-10       Impact factor: 3.481

6.  A comparison of computer-controlled versus manual on-line patient setup adjustment.

Authors:  Kristy K Brock; Daniel L McShan; James M Balter
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

7.  Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system.

Authors:  Timothy Fox; Calvin Huntzinger; Peter Johnstone; Tomi Ogunleye; Eric Elder
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.