Literature DB >> 16185038

[Prospective screw misplacement analysis after conventional and navigated pedicle screw implantation].

K Seller1, A Wild, L Urselmann, R Krauspe.   

Abstract

BACKGROUND: [corrected] The aim of this prospective study was (1) to evaluate the accuracy of pedicle screw placement using Computer - Assisted Orthopedic - Surgery (CAOS) in comparison to conventionelly image intensifier controlled pedicle screw instrumentation, (2) to compare our results with data from literature and (3) report our experiences with this technique. PATIENTS AND METHODS: Between 11/00 and 11/01 sixteen patients planned for spine surgery were subsequently recruited. Pedicle screw instrumentation was done in each patient as well with computer aided surgery (CAOS, SurgiGate-System, Medivision, Stratec Medical, Swiss) as also with image intensifier control, allowing for intraindividual comparison. Evaluation of pedicle screw placement was carried out with postoperative computed tomography (CT) or magnetic resonance imaging (MRI).
RESULTS: 33 of altogether 36 pedicle screws inserted with Computer-Assistance (CAOS) were correctly placed (91,7%), however only 17 of altogether 24 pedicle screws inserted under image intensifier control (70,8%). The difference of frequency of screw misplacement between Computer-aided and image intensifier controlled instrumentation was statistically significant (p<0.05; chi-square test).
CONCLUSION: Computer assisted surgery reduces significantly the misplacement rate of pedicle screws and remains for experienced spine surgeons an important support in the operative treatment of complex spinal deformities in future. Additionally it can be expected that Computer-Navigation will also spread out in the field of minimal-invasive spinal surgery, e.g. the kyphoplasty. The use of this technique supports beside the medical-technical knowledge an improved three-dimensional orientation in the education of spine surgeons.

Entities:  

Mesh:

Year:  2005        PMID: 16185038     DOI: 10.1515/BMT.2005.043

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411


  6 in total

Review 1.  Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies.

Authors:  Nai-Feng Tian; Qi-Shan Huang; Ping Zhou; Yang Zhou; Rui-Kai Wu; Yi Lou; Hua-Zi Xu
Journal:  Eur Spine J       Date:  2010-09-23       Impact factor: 3.134

Review 2.  Computer navigation versus fluoroscopy-guided navigation for thoracic pedicle screw placement: a meta-analysis.

Authors:  Xiao-Tong Meng; Xiao-Fei Guan; Hai-Long Zhang; Shi-Sheng He
Journal:  Neurosurg Rev       Date:  2015-12-19       Impact factor: 3.042

3.  Image-guided pedicle screw insertion accuracy: a meta-analysis.

Authors:  Nai-Feng Tian; Hua-Zi Xu
Journal:  Int Orthop       Date:  2009-05-08       Impact factor: 3.075

Review 4.  Functional outcome of computer-assisted spinal pedicle screw placement: a systematic review and meta-analysis of 23 studies including 5,992 pedicle screws.

Authors:  Rajeev Verma; Sonal Krishan; Kurt Haendlmayer; A Mohsen
Journal:  Eur Spine J       Date:  2010-01-06       Impact factor: 3.134

5.  Pedicle screw piercer with warning device - A technique to increase accuracy of pedicle screw placement: A cadaveric study.

Authors:  Lin Bin; He Yong; Xu Yang; Zhang Bi; Sha Mo; Guo Zhi-Min
Journal:  Indian J Orthop       Date:  2014-11       Impact factor: 1.251

6.  Effective dose of radiation per screw in surgery of adolescent idiopathic scoliosis: matched pair analysis of 293 pedicle screws inserted using three different techniques.

Authors:  Markus Rafael Konieczny; Johannes Boos; Andrea Steuwe; Christoph Schleich; Max Prost; Rüdiger Krauspe
Journal:  J Child Orthop       Date:  2020-12-01       Impact factor: 1.548

  6 in total