Literature DB >> 16170808

Comparison of manual and automated ELISA methods for serum ferritin analysis.

Fabian Rohner1, Christophe Zeder, Michael B Zimmermann, Richard F Hurrell.   

Abstract

Serum ferritin concentration is a sensitive measure of body iron stores. The aim of this study was to compare the performance of two commercially available enzyme-linked immunoassays (ELISAs) for serum ferritin: a widely used manual assay kit (Spectro Ferritin MT), and a new fully automated assay (Immulite). We analyzed serum samples from Moroccan school-aged children (n=51) from a rural area with a high prevalence of iron deficiency anemia (IDA). Four replicates of each sample were analyzed using both assays. For the manual method, the interassay repeatability was 24%, 22%, and 11%, and intraassay precision was 18.3%, 9.2%, and 9.1% at increasing serum ferritin concentrations. Using the automated assay, the interassay repeatability was 7%, 6%, and 6%, and intraassay precision was 1.5%, 5.4%, and 5.5% at increasing serum ferritin concentrations. The two assays were well correlated (y=1.16x+1.83; r=0.98). However, the limits of agreement (LOAs) were wide, particularly at low concentrations. A comparison of the assay results with recommended cutoffs for serum ferritin generated sharply different estimates of the prevalence of iron deficiency (ID) in the sample. We conclude that the automated assay has several potential advantages compared to the manual method, including better precision, less operator dependence, and faster sample through-put. (c) 2005 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16170808      PMCID: PMC6808104          DOI: 10.1002/jcla.20077

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  6 in total

1.  Measurement of ferritin levels: comparison of a commercial IRMA to an in-house ELISA method.

Authors:  S Zemelka; H K Biesalski
Journal:  J Immunoassay Immunochem       Date:  2001

2.  Clinical utility of the soluble transferrin receptor and comparison with serum ferritin in several populations.

Authors:  A E Mast; M A Blinder; A M Gronowski; C Chumley; M G Scott
Journal:  Clin Chem       Date:  1998-01       Impact factor: 8.327

3.  Treatment of iron deficiency in goitrous children improves the efficacy of iodized salt in Côte d'Ivoire.

Authors:  Sonja Y Hess; Michael B Zimmermann; Pierre Adou; Toni Torresani; Richard F Hurrell
Journal:  Am J Clin Nutr       Date:  2002-04       Impact factor: 7.045

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 5.  Defining optimal body iron.

Authors:  J D Cook
Journal:  Proc Nutr Soc       Date:  1999-05       Impact factor: 6.297

6.  Dual fortification of salt with iodine and microencapsulated iron: a randomized, double-blind, controlled trial in Moroccan schoolchildren.

Authors:  Michael B Zimmermann; Christophe Zeder; Noureddine Chaouki; Amina Saad; Toni Torresani; Richard F Hurrell
Journal:  Am J Clin Nutr       Date:  2003-02       Impact factor: 7.045

  6 in total
  4 in total

1.  Somatostatin receptor immunohistochemistry in neuroendocrine tumors: comparison between manual and automated evaluation.

Authors:  Daniel Kaemmerer; Maria Athelogou; Amelie Lupp; Isabell Lenhardt; Stefan Schulz; Peter Luisa; Merten Hommann; Vikas Prasad; Gerd Binnig; Richard Paul Baum
Journal:  Int J Clin Exp Pathol       Date:  2014-07-15

2.  Co-assessment of iron, vitamin A and growth status to investigate anemia in preschool children in suburb Chongqing, China.

Authors:  Ke Chen; Xuan Zhang; Ting-Yu Li; Li Chen; Ping Qu; You-Xue Liu
Journal:  World J Pediatr       Date:  2009-11-13       Impact factor: 2.764

3.  A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods.

Authors:  Lotfi S Bin Dahman; Khalid M Sumaily; Essa M Sabi; Mohammed A Hassan; Abeer M Bin Thalab; Asrar S Sayad; Saleh M Bin Kolaib; Fatima M Alhadhrmi
Journal:  Diagnostics (Basel)       Date:  2022-01-27

Review 4.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.