Interface reaction at dental implants inserted in condensed bone.

André Büchter1, Johannes Kleinheinz, Hans Peter Wiesmann, Mythili Jayaranan, Ulrich Joos, Ulrich Meyer.   

Abstract

PURPOSE: The influence of the osteotome technique on the interface reaction of cylinder implants (SLA, ITI) was compared with the interface reaction of conventional implant insertion in an animal model.
MATERIAL AND METHODS: A total of 64 implants were placed in the cranial and caudal tibia of 8 Göttinger minipigs. The implant site was prepared either by a conventional technique with drills (control group A) or by the osteotome technique (experimental group B). Bone tissue responses were evaluated by histomorphometry, fluorescence microscopy and scanning electron microscopy after 7 and 28 days of osseointegration.
RESULTS: The average initial (7 days) bone-to-implant contact ratio was not statistically significantly different for the osteotome technique (35.88+/-2.94%) than for the control group (43.78+/-3.39%, P<0.095). After 28 days, the bone-to-implant contact ratio became statistically significantly higher when implants were inserted by conventional preparation (44.81+/-3.07% (group B), 63.47+/-4.87% (group A), P=0.003). Whereas fluorescence and immunhistologic examination revealed new bone formation with osteocalcin deposition directly at the implant surface in both groups, the extent of direct bone/implant contact was enhanced in conventionally prepared implant sites. SEM analysis confirmed an intimate bone to implant bond without fibrous tissue formation in places of direct contact at an ultrastructured level.
CONCLUSION: Implant placement in conventionally prepared implantation sites is accompanied by an improved interface formation at an early stage of implantation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16164455     DOI: 10.1111/j.1600-0501.2005.01111.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  8 in total

1.  Clinical indications, advantages and limits of the expansion-condensing osteotomes technique for the creation of implant bed.

Authors:  G DE Vico; M Bonino; D Spinelli; A Pozzi; A Barlattani
Journal:  Oral Implantol (Rome)       Date:  2009-12-10

2.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

3.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

Review 4.  Does the Implant Surgical Technique Affect the Primary and/or Secondary Stability of Dental Implants? A Systematic Review.

Authors:  Rola Muhammed Shadid; Nasrin Rushdi Sadaqah; Sahar Abdo Othman
Journal:  Int J Dent       Date:  2014-07-07

5.  Customized SmartPeg for measurement of resonance frequency of mini dental implants.

Authors:  Jagjit Singh Dhaliwal; Rubens F Albuquerque; Ali Fakhry; Sukhbir Kaur; Jocelyne S Feine
Journal:  Int J Implant Dent       Date:  2017-02-01

6.  The developing patterns of calibrated implant stability quotients of posterior implants.

Authors:  Chih-Chun Mei; Yen-Ting Lin; Chung-Chieh Chang; Alex Hong; Hsiang-Hsi Hong
Journal:  J Dent Sci       Date:  2021-05-06       Impact factor: 2.080

7.  Investigation of biomaterials by human epithelial gingiva cells: an in vitro study.

Authors:  Jörg Neunzehn; Beate Lüttenberg; Hans-Peter Wiesmann
Journal:  Head Face Med       Date:  2012-12-15       Impact factor: 2.151

8.  Correlation between Insertion Torque and Implant Stability Quotient in Tapered Implants with Knife-Edge Thread Design.

Authors:  Domenico Baldi; Teresa Lombardi; Jacopo Colombo; Gabriele Cervino; Giuseppe Perinetti; Roberto Di Lenarda; Claudio Stacchi
Journal:  Biomed Res Int       Date:  2018-05-15       Impact factor: 3.411

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.