DPSS yellow-green 561-nm lasers for improved fluorochrome detection by flow cytometry.

William Telford1, Matilde Murga, Teresa Hawley, Robert Hawley, Beverly Packard, Akira Komoriya, Fred Haas, Charles Hubert.   

Abstract

INTRODUCTION: Blue-green 488-nm laser sources are widespread in flow cytometry but suffer some drawbacks for cell analysis, including their excitation of endogenous proteins (resulting in high cellular autofluorescence) and their less-than-optimal coincidence with the excitation maxima of commonly used fluorochromes, including the phycoerythrins (PE). Longer wavelength lasers such as green helium-neons and, more recently, diode-pumped solid state (DPSS) 532-nm sources have previously been employed to overcome these difficulties and improve overall sensitivity for PE. In this study, we evaluate an even longer wavelength DPSS 561-nm for its ability to improve PE and DsRed fluorescent protein detection sensitivity.
METHODS: A DPSS 561-nm laser emitting at 10 mW was mounted onto a BD LSR II. Mouse thymoma cells labeled with cell surface marker antibodies conjugated to the R- and B-forms of PE were analyzed and compared with conventional 488-nm excitation using the same bandpass filters and signal travel distances. A similar analysis was carried out with cell lines expressing the red fluorescent protein DsRed, several green-yellow excited low molecular weight fluorochromes, and a rhodamine-based caspase substrate. Additionally, cells labeled with PE and co-labeled with fluorescein or simultaneously expressing green fluorescent protein (GFP) were analyzed to determine if PE excitation at 561 nm with simultaneous fluorescein/GFP detection was feasible.
RESULTS: The DPSS 561-nm laser gave a several-fold improvement in the fluorochrome to autofluorescence ratios between PE-labeled cells and unlabeled controls. Analysis of cells expressing the fluorescent protein DsRed with the DPSS 561-nm source gave a 6-7-fold improvement in sensitivity over 488-nm excitation, and gave excellent excitation of yellow-green excited fluorochromes and rhodamine-based physiological probes. Yellow-green laser light also caused virtually no impingement on the spatially separated fluorescein/GFP detector, a significant problem with green laser sources, and also allowed simultaneous analysis of GFP and PE with virtually no signal overlap or requirement for color compensation.
CONCLUSIONS: DPSS 561-nm laser excitation gave significantly improved sensitivity for both PE-labeled and DsRed expressing cells, with little contamination of a typical fluorescein/GFP detector.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16163703     DOI: 10.1002/cyto.a.20182

Source DB:  PubMed          Journal:  Cytometry A        ISSN: 1552-4922            Impact factor:   4.355


  15 in total

1.  Diode-pumped solid-state laser for bonding orthodontic brackets: effect of light intensity and light-curing time.

Authors:  Young-Oh Kim; Soo-Byung Park; Woo-Sung Son; Ching-Chang Ko; Franklin García-Godoy; Hyung-Il Kim; Yong Hoon Kwon
Journal:  Lasers Med Sci       Date:  2010-06-08       Impact factor: 3.161

2.  The applicability of DPSS laser for light curing of composite resins.

Authors:  Yong Hoon Kwon; Chang-Min Jang; Dong-Hee Shin; Hyo-Joung Seol; Hyung-Il Kim
Journal:  Lasers Med Sci       Date:  2007-09-28       Impact factor: 3.161

3.  Effect of diode-pumped solid state laser on polymerization shrinkage and color change in composite resins.

Authors:  Yun-Hoa Jung; Bong-Hae Cho; Kyung-Soo Nah; Hyung-Il Kim; Yong Hoon Kwon
Journal:  Lasers Med Sci       Date:  2009-02-10       Impact factor: 3.161

4.  Solid state yellow and orange lasers for flow cytometry.

Authors:  Veena Kapoor; Vladimir Karpov; Claudette Linton; Fedor V Subach; Vladislav V Verkhusha; William G Telford
Journal:  Cytometry A       Date:  2008-06       Impact factor: 4.355

5.  Microhardness and polymerization shrinkage of flowable resins that are light cured using a blue laser.

Authors:  Chang-Mo Jeong; Young-Joon Heo; Young-Chan Jeon; Hyung-Il Kim; Yong Hoon Kwon
Journal:  Lasers Med Sci       Date:  2011-07-24       Impact factor: 3.161

6.  Effect of 457 nm diode-pumped solid state laser on the polymerization composite resins: microhardness, cross-link density, and polymerization shrinkage.

Authors:  Sung-Ae Son; Jeong-Kil Park; Kyoung-Hwa Jung; Ching-Chang Ko; Chang-Mo Jeong; Yong Hoon Kwon
Journal:  Photomed Laser Surg       Date:  2014-12-30       Impact factor: 2.796

7.  Effect of a DPSS laser on the shear bond strength of ceramic brackets with different base designs.

Authors:  Mi-Gyoung Park; Jung-Hoon Ro; Jeong-Kil Park; Ching-Chang Ko; Yong Hoon Kwon
Journal:  Lasers Med Sci       Date:  2012-11-08       Impact factor: 3.161

8.  Supercontinuum white light lasers for flow cytometry.

Authors:  William G Telford; Fedor V Subach; Vladislav V Verkhusha
Journal:  Cytometry A       Date:  2009-05       Impact factor: 4.355

9.  Systems biological analysis of epidermal growth factor receptor internalization dynamics for altered receptor levels.

Authors:  Hannah Schmidt-Glenewinkel; Eileen Reinz; Roland Eils; Nathan R Brady
Journal:  J Biol Chem       Date:  2009-03-17       Impact factor: 5.157

10.  Green fiber lasers: an alternative to traditional DPSS green lasers for flow cytometry.

Authors:  William G Telford; Sergey A Babin; Serge V Khorev; Stephen H Rowe
Journal:  Cytometry A       Date:  2009-12       Impact factor: 4.355

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.