| Literature DB >> 16161124 |
Gulsen Bayraktar1, Bora Guvener, Canan Bural, Yagiz Uresin.
Abstract
This study compared the influence of different polymerization methods (heat, auto-, and microwave energy), different curing processes (in the case of heat- and autopolymerized specimens), and length of storage of the polymerized specimens in distilled water at 37 degrees C on the residual methyl methacrylate (MMA) content in dental acrylic resin specimens. Residual MMA of 120 resin specimens were measured using high-performance liquid chromatography. For the heat-polymerized resins, the lowest residual MMA content was obtained when they were given a long-term terminal boil and then stored in the distilled water for at least 1 day. For the autopolymerized resins, the lowest residual MMA content was obtained when they were additionally cured in water at 60 degrees C and then stored in the distilled water at least 1 day. For the microwave-polymerized resins, the lowest residual MMA content was obtained when they were stored in the distilled water at least 1 month. The lowest overall residual MMA content was obtained from heat-polymerized specimens that were given a long-term terminal boil cure and then stored in the distilled water at least 1 day. Different polymerization methods and curing processes have different effects on residual MMA content. It is thus shown that storing a dental acrylic resin specimen in distilled water at 37 degrees C is a simple but effective method of reducing its residual MMA content. Copyright 2005 Wiley Periodicals, Inc.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16161124 DOI: 10.1002/jbm.b.30377
Source DB: PubMed Journal: J Biomed Mater Res B Appl Biomater ISSN: 1552-4973 Impact factor: 3.368