Comparison of the Lactate Pro and Analox GM7 blood lactate analysers.

K A van Someren1, G Howatson, D Nunan, R Thatcher, R Shave.   

Abstract

The purpose of this study was firstly, to determine the level of agreement between the Lactate Pro and Analox GM7 for the measurement of blood lactate, and secondly, to examine whether these analysers may be used interchangeably to identify lactate parameters routinely used in the physiological assessment of athletes. Twenty well-trained male cyclists performed an incremental cycle ergometry test; duplicate blood samples were taken simultaneously throughout the test for lactate determination using the two analysers. Power output and heart rate at LT, LT1, 2 mmol.L-1, and 4 mmol.L-1 were calculated from the lactate values obtained from the two analysers. There was a strong linear relationship between the two analysers (Lactate Pro=1.4541xAnalox GM7-0.1287; R2=0.969, p<0.001; SEE=0.704) though the Lactate Pro overestimated blood lactate when compared to the Analox GM7; 95% ratio limits of agreement (Lactate Pro/Analox GM7) were 1.40x//1.35. Differences were found between analysers for power output and heart rate at LT1 (p<0.01), 2 mmol.L-1 (p<0.001), and 4 mmol.L-1 (p<0.001); no differences were found between analysers for power output or heart rate at LT. In conclusion, the average difference in blood lactate concentration measured by the Lactate Pro and the Analox GM7 was 40%, with 95% of measures differing by between 4% and 89%. This poor level of agreement and the significant differences between the two analysers in power output and heart rate associated with a number of lactate parameters indicate that these analysers should not be used interchangeably.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16158371     DOI: 10.1055/s-2004-830337

Source DB:  PubMed          Journal:  Int J Sports Med        ISSN: 0172-4622            Impact factor:   3.118


  4 in total

1.  Reliability and accuracy of six hand-held blood lactate analysers.

Authors:  Jacinta M Bonaventura; Ken Sharpe; Emma Knight; Kate L Fuller; Rebecca K Tanner; Christopher J Gore
Journal:  J Sports Sci Med       Date:  2015-03-01       Impact factor: 2.988

2.  Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus.

Authors:  Rebecca K Tanner; Kate L Fuller; Megan L R Ross
Journal:  Eur J Appl Physiol       Date:  2010-02-10       Impact factor: 3.078

Review 3.  Lactate threshold concepts: how valid are they?

Authors:  Oliver Faude; Wilfried Kindermann; Tim Meyer
Journal:  Sports Med       Date:  2009       Impact factor: 11.136

4.  Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer.

Authors:  Carlo Baldari; Valerio Bonavolontà; Gian Pietro Emerenziani; Maria Chiara Gallotta; Antonio José Silva; Laura Guidetti
Journal:  Eur J Appl Physiol       Date:  2009-06-14       Impact factor: 3.078

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.