Literature DB >> 16121595

Design and application of an assessment protocol for electromagnetic tracking systems.

Johann B Hummel1, Michael R Bax, Michael L Figl, Yan Kang, Calvin Maurer, Wolfgang W Birkfellner, Helmar Bergmann, Ramin Shahidi.   

Abstract

This paper defines a simple protocol for competitive and quantified evaluation of electromagnetic tracking systems such as the NDI Aurora (A) and Ascension microBIRD with dipole transmitter (B). It establishes new methods and a new phantom design which assesses the reproducibility and allows comparability with different tracking systems in a consistent environment. A machined base plate was designed and manufactured in which a 50 mm grid of holes was precisely drilled for position measurements. In the center a circle of 32 equispaced holes enables the accurate measurement of rotation. The sensors can be clamped in a small mount which fits into pairs of grid holes on the base plate. Relative positional/orientational errors are found by subtracting the known distances/ rotations between the machined locations from the differences of the mean observed positions/ rotation. To measure the influence of metallic objects we inserted rods made of steel (SST 303, SST 416), aluminum, and bronze into the sensitive volume between sensor and emitter. We calculated the fiducial registration error and fiducial location error with a standard stylus calibration for both tracking systems and assessed two different methods of stylus calibration. The positional jitter amounted to 0.14 mm(A) and 0.08 mm(B). A relative positional error of 0.96 mm +/- 0.68 mm, range -0.06 mm; 2.23 mm(A) and 1.14 mm +/- 0.78 mm, range -3.72 mm; 1.57 mm(B) for a given distance of 50 mm was found. The relative rotation error was found to be 0.51 degrees (A)/0.04 degrees (B). The most relevant distortion caused by metallic objects results from SST 416. The maximum error 4.2 mm(A)/ > or = 100 mm(B) occurs when the rod is close to the sensor(20 mm). While (B) is more sensitive with respect to metallic objects, (A) is less accurate concerning orientation measurements. (B) showed a systematic error when distances are calculated.

Entities:  

Mesh:

Year:  2005        PMID: 16121595     DOI: 10.1118/1.1944327

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Electromagnetic tracking for US-guided interventions: standardized assessment of a new compact field generator.

Authors:  A M Franz; K März; J Hummel; W Birkfellner; R Bendl; S Delorme; H-P Schlemmer; H-P Meinzer; L Maier-Hein
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-05-24       Impact factor: 2.924

2.  Augmented reality haptic (ARH): an approach of electromagnetic tracking in minimally invasive surgery.

Authors:  J B Pagador; L F Sánchez; J A Sánchez; P Bustos; J Moreno; F M Sánchez-Margallo
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-07-02       Impact factor: 2.924

3.  Novel 3-D laparoscopic magnetic ultrasound image guidance for lesion targeting.

Authors:  David Sindram; Iain H McKillop; John B Martinie; David A Iannitti
Journal:  HPB (Oxford)       Date:  2010-12       Impact factor: 3.647

4.  Electromagnetic tracking in the clinical environment.

Authors:  Ziv Yaniv; Emmanuel Wilson; David Lindisch; Kevin Cleary
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

5.  Distorter Characterisation Using Mutual Inductance in Electromagnetic Tracking.

Authors:  Herman Alexander Jaeger; Pádraig Cantillon-Murphy
Journal:  Sensors (Basel)       Date:  2018-09-12       Impact factor: 3.576

Review 6.  Navigated laparoscopic ultrasound in abdominal soft tissue surgery: technological overview and perspectives.

Authors:  Thomas Langø; Sinara Vijayan; Anna Rethy; Cecilie Våpenstad; Ole Vegard Solberg; Ronald Mårvik; Gjermund Johnsen; Toril N Hernes
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-09-03       Impact factor: 2.924

7.  An analysis of tracking error in image-guided neurosurgery.

Authors:  Ian J Gerard; D Louis Collins
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-01-04       Impact factor: 2.924

8.  Electromagnetically tracked personalized templates for surgical navigation.

Authors:  Andrew W L Dickinson; Michelle L Zec; David R Pichora; Brian J Rasquinha; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-22       Impact factor: 2.924

9.  First clinical use of the EchoTrack guidance approach for radiofrequency ablation of thyroid gland nodules.

Authors:  Alfred Michael Franz; Alexander Seitel; Nasrin Bopp; Christian Erbelding; Dominique Cheray; Stefan Delorme; Frank Grünwald; Hüdayi Korkusuz; Lena Maier-Hein
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-22       Impact factor: 2.924

10.  Anser EMT: the first open-source electromagnetic tracking platform for image-guided interventions.

Authors:  Herman Alexander Jaeger; Alfred Michael Franz; Kilian O'Donoghue; Alexander Seitel; Fabian Trauzettel; Lena Maier-Hein; Pádraig Cantillon-Murphy
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-29       Impact factor: 2.924

View more