Literature DB >> 15988111

Valve dysfunction of the cloth-covered Starr-Edwards ball valve.

Motomi Shiono1, Akira Sezai, Mitsumasa Hata, Mitsuru Iida, Nanao Negishi, Yukiyasu Sezai.   

Abstract

BACKGROUND: Between June 1968 and March 1977, Starr-Edwards cloth-covered ball valves were used for valve replacement on a routine basis. METHODS AND
RESULTS: Among the 66 operative survivors who underwent an isolated aortic or mitral valve replacement, 20 patients required reoperation 22 times because of valve dysfunction, thromboembolic complication, paravalvular leakage, hemolytic anemia, and/or prosthetic valve endocarditis. Reoperation was performed at a mean of 15.9+/-9.8 years after initial replacement. Excised valves were examined and reoperation after initial operation was reviewed. Operative mortality was 10.0%. Freedom from reoperation for aortic valve replacement and mitral valve replacement was 56.2% at 34 years and 61.0% at 37 years after initial operation, respectively. Cloth wear or pannus formation were observed in all excised prostheses. Orifice cloth was more markedly worn in mitral valves than in aortic valves, particularly in mitral valves of more than 20 years old. Pannus overgrowth contributed to valve regurgitation in the older valves.
CONCLUSIONS: Early diagnosis of valve dysfunction and reoperation are recommended as soon as symptoms appear.

Entities:  

Mesh:

Year:  2005        PMID: 15988111     DOI: 10.1253/circj.69.844

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  3 in total

1.  Reoperation on a Starr-Edwards ball valve without structural deterioration.

Authors:  Yukihiro Hayatsu; Takeshi Saito; Osamu Adachi; Kiichiro Kumagai; Masatoshi Akiyama; Naotaka Motoyoshi; Shunsuke Kawamoto; Yoshikatsu Saiki
Journal:  Gen Thorac Cardiovasc Surg       Date:  2012-05-29

2.  A case of the Starr-Edwards ball valve (Model 6120) in the mitral position for 45 years.

Authors:  Shuhei Azuma; Masafumi Morita; Yasuhiro Yoshii; Shigetoshi Mieno
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-07-19

3.  Use of the Bentall method for replacement of an aortic Starr-Edwards ball valve prosthesis implanted 32 years previously.

Authors:  Daisuke Yotsumoto; Yoshifumi Iguro; Takayuki Ueno; Kazuhisa Matsumoto; Ryuzo Sakata
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.