Polyp size at CT colonography after electronic subtraction cleansing in an anthropomorphic colon phantom.

Michael E Zalis1, James J Perumpillichira, John Y Kim, Chiara Del Frate, Cordula Magee, Peter F Hahn.   

Abstract

PURPOSE: To evaluate the effect of various bowel contrast material concentrations and subtraction software on size measurements of well-defined polyp lesions in a colon phantom at CT colonography.
MATERIALS AND METHODS: Repeated scanning and a precise reference standard required the use of a colon phantom in which 21 polyps were randomly distributed. Two readers who had each reviewed computed tomographic (CT) colonographic images from more than 100 cases evaluated polyp size on images obtained when the phantom was partially filled with varying concentrations of contrast material, scanned by using CT colonography, and subjected to electronic subtraction cleansing. The single largest dimension was recorded for each reader for a randomized series of polyps. These measurements were compared with a reference standard that was based on a combination of the manufacturer's polyp size specifications and the subsequent verification of these sizes by an independent consensus panel. Six weeks after initial observations, readers evaluated images of the phantom scanned without the presence of contrast material. Polyp size estimations for the two readers for each series were compared with the reference standard to obtain a mean absolute measurement error for each reader for each series. Data for each reader were compared by using a nonparametric Kruskal-Wallis analysis of variance test. A pair-wise comparison of the experimental and control series was then performed by using the Dunn post hoc test.
RESULTS: Contrast material dilutions resulting in an average attenuation of less than 500 HU resulted in complete subtraction and the absence of streak artifacts. There was no statistically significant difference between the average measurement error for contrast attenuations between 300 and 500 HU when compared with that of control. Streak artifact was noticeable for the highest dilution (mean, 840 HU). No statistically significant differences were observed for series in which cleansing software was used in the absence of bowel contrast material.
CONCLUSION: The combination of electronic cleansing and bowel contrast enhancement in the range of 300-500 HU results in no substantial change in readers' estimations of polyp size at CT colonography. Copyright RSNA, 2005

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15987967     DOI: 10.1148/radiol.2361040231

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  14 in total

Review 1.  Polyp size measurement at CT colonography: what do we know and what do we need to know?

Authors:  Ronald M Summers
Journal:  Radiology       Date:  2010-06       Impact factor: 11.105

2.  Fully automated three-dimensional detection of polyps in fecal-tagging CT colonography.

Authors:  Janne Näppi; Hiroyuki Yoshida
Journal:  Acad Radiol       Date:  2007-03       Impact factor: 3.173

3.  Adaptive correction of the pseudo-enhancement of CT attenuation for fecal-tagging CT colonography.

Authors:  Janne Näppi; Hiroyuki Yoshida
Journal:  Med Image Anal       Date:  2008-01-26       Impact factor: 8.545

4.  [CT colonography: techniques of visualization and findings].

Authors:  J Wessling; W Heindel
Journal:  Radiologe       Date:  2008-02       Impact factor: 0.635

5.  Structure-analysis method for electronic cleansing in cathartic and noncathartic CT colonography.

Authors:  Wenli Cai; Michael E Zalis; Janne Näppi; Gordon J Harris; Hiroyuki Yoshida
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

6.  Nonlinear regression-based method for pseudoenhancement correction in CT colonography.

Authors:  Baigalmaa Tsagaan; Janne Näppi; Hiroyuki Yoshida
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

7.  Informatics in radiology: Electronic cleansing for noncathartic CT colonography: a structure-analysis scheme.

Authors:  Wenli Cai; Hiroyuki Yoshida; Michael E Zalis; Janne Johannes Näppi; Gordon J Harris
Journal:  Radiographics       Date:  2010-03-10       Impact factor: 5.333

8.  Intra-individual comparison of magnesium citrate and sodium phosphate for bowel preparation at CT colonography: automated volumetric analysis of residual fluid for quality assessment.

Authors:  P Bannas; J Bakke; A Munoz del Rio; P J Pickhardt
Journal:  Clin Radiol       Date:  2014-09-18       Impact factor: 2.350

9.  Mosaic decomposition: an electronic cleansing method for inhomogeneously tagged regions in noncathartic CT colonography.

Authors:  Wenli Cai; June-Goo Lee; Michael E Zalis; Hiroyuki Yoshida
Journal:  IEEE Trans Med Imaging       Date:  2010-10-14       Impact factor: 10.048

10.  Scale-based scatter correction for computer-aided polyp detection in CT colonography.

Authors:  Jiamin Liu; Jianhua Yao; Ronald M Summers
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.