T Kriebel1, W Ruschewski, T Paul.
Abstract
In infants and small children, ICD implantation is a challenge due to technical limitations and a significant number of complications. This report describes ICD implantation in a 6-month-old infant (body weight 5.5 kg). A completely extracardiac defibrillation system was implanted using a transvenous lead subcutaneously in the back below the left scapula as the defibrillation electrode and an active-can device in the right upper abdomen. Defibrillation threshold of implantation was < or =10 J. During the follow-up of 3 months, 8 adequate ICD discharges were noted. The technique described seems feasible to facilitate ICD implantation in small infants.Entities:
Mesh:
Year: 2005 PMID: 15940443 DOI: 10.1007/s00392-005-0236-z