Literature DB >> 15920420

Cyclic loading of sacroiliac screws in Tile C pelvic fractures.

C M A van Zwienen1, Eric W van den Bosch, G A Hoek van Dijke, C J Snijders, Arie B van Vugt.   

Abstract

BACKGROUND: To investigate the stiffness and strength of completely unstable pelvic fractures fixated both anteriorly and posteriorly under cyclic loading conditions, the authors conducted a randomized, comparative, cadaveric study.
METHODS: In 12 specimens, a Tile C1 pelvic fracture was created. The authors compared the intact situation to anterior plate fixation combined with one or two sacroiliac screws. In 2,000 measurements, each pelvis was loaded with a maximum of 400 N. The translation and rotation stiffness of the fixations were measured using a three-dimensional video system. Furthermore, the load to failure and the number of cycles before failure were determined.
RESULTS: Both translation and rotation stiffness of the intact pelvis were superior to the fixated pelvis. No difference in stiffness was found between the techniques with one or two sacroiliac screws. However, a significantly higher load to failure and significantly more loading cycles before failure could be achieved using two sacroiliac screws compared with one screw.
CONCLUSION: Although the combination of anterior plate fixation combined with two sacroiliac screws is not as stable as the intact pelvis, in this study, embalmed aged pelves could be loaded repeatedly with physiologic forces. Given the fact that the average trauma patient is younger and given the fact that the quality (or grip) of the fixation was a significant covariable for longer endurance of the fixation, this suggests that direct postoperative weight bearing could be possible if these results are confirmed in further research.

Entities:  

Mesh:

Year:  2005        PMID: 15920420     DOI: 10.1097/01.ta.0000158515.58494.11

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  15 in total

1.  Analysis of sacro-iliac joint screw fixation: does quality of reduction and screw orientation influence joint stability? A biomechanical study.

Authors:  Gaston Camino Willhuber; Ivan Zderic; Florian Gras; Dieter Wahl; Carlos Sancineto; Jorge Barla; Markus Windolf; Robert Geoff Richards; Boyko Gueorguiev
Journal:  Int Orthop       Date:  2015-10-05       Impact factor: 3.075

2.  Biomechanical comparison of three types of internal fixation in a type C zone II pelvic fracture model.

Authors:  Tao Wu; Wei Chen; Xu Li; Qi Zhang; Hong-Zhi Lv; Ying-Ze Zhang
Journal:  Int J Clin Exp Med       Date:  2015-02-15

3.  Triangular Titanium Implants for Minimally Invasive Sacroiliac Joint Fusion: 2-Year Follow-Up from a Prospective Multicenter Trial.

Authors:  Bradley S Duhon; Fabien Bitan; Harry Lockstadt; Don Kovalsky; Daniel Cher; Travis Hillen
Journal:  Int J Spine Surg       Date:  2016-04-20

4.  When do anterior external or internal fixators provide additional stability in an unstable (Tile C) pelvic fracture? A biomechanical study.

Authors:  E Mcdonald; A A Theologis; P Horst; U Kandemir; M Pekmezci
Journal:  Eur J Trauma Emerg Surg       Date:  2014-12-03       Impact factor: 3.693

5.  Symphyseal internal rod fixation versus standard plate fixation for open book pelvic ring injuries: a biomechanical study.

Authors:  G Osterhoff; S Tiziani; C Hafner; S J Ferguson; H-P Simmen; C M L Werner
Journal:  Eur J Trauma Emerg Surg       Date:  2015-04-17       Impact factor: 3.693

6.  Anterior internal fixator versus a femoral distractor and external fixation for sacroiliac joint compression and single stance gait testing: a mechanical study in synthetic bone.

Authors:  Jonathan M Vigdorchik; Amanda O Esquivel; Xin Jin; King H Yang; Rahul Vaidya
Journal:  Int Orthop       Date:  2013-05-04       Impact factor: 3.075

7.  Therapeutic effects of minimally invasive adjustable and locking compression plate for unstable pelvic fractures via posterior approach.

Authors:  Tao Wu; Wei Chen; Qi Zhang; Xu Li; Hong-Zhi Lv; Guang Yang; Ying-Ze Zhang
Journal:  Int J Clin Exp Med       Date:  2015-01-15

Review 8.  [Biomechanical principles for treatment of osteoporotic fractures of the pelvis].

Authors:  A Gänsslen
Journal:  Unfallchirurg       Date:  2010-04       Impact factor: 1.000

9.  Posterior tension band plate osteosynthesis for unstable sacral fractures: A preliminary study.

Authors:  Veliyaveettil Muhamed Khaleel; Narendran Pushpasekaran; Nithin Prabhu; Alagu Pandiyan; George Mathew Koshy
Journal:  J Clin Orthop Trauma       Date:  2019-05-24

10.  Biomechanical Comparison of Two Kinds of Internal Fixation in a Type C Zone II Pelvic Fracture Model.

Authors:  Tao Wu; Wei Chen; Qi Zhang; Zhan-Le Zheng; Hong-Zhi Lyu; Yun-Wei Cui; Xiao-Dong Cheng; Ying-Ze Zhang; Yan-Jiang Yang
Journal:  Chin Med J (Engl)       Date:  2015-09-05       Impact factor: 2.628