Computational study of fluid mechanical disturbance induced by endovascular stents.

Taewon Seo1, Levanto G Schachter, Abdul I Barakat.   

Abstract

Arterial restenosis following stent deployment may be influenced by the local flow environment within and around the stent. We have used computational fluid dynamics to investigate the flow field in the vicinity of model stents positioned within straight and curved vessels. Our simulations have revealed the presence of flow separation and recirculation immediately downstream of stents. In steady flow within straight vessels, the extent of flow disturbance downstream of the stent increases with both Reynolds number and stent wire thickness but is relatively insensitive to stent interwire spacing. In curved vessels, flow disturbance downstream of the stent occurs along both the inner and outer vessel walls with the extent of disturbance dependent on the angle of vessel curvature. In pulsatile flow, the regions of flow disturbance periodically increase and decrease in size. Non-Newtonian fluid properties lead to a modest reduction in flow disturbance downstream of the stent. In more realistic stent geometries such as stents modeled as spirals or as intertwined rings, the nature of stent-induced flow disturbance is exquisitely sensitive to stent design. These results provide an understanding of the flow physics in the vicinity of stents and suggest strategies for stent design optimization to minimize flow disturbance.

Entities:  

Mesh:

Year:  2005        PMID: 15909650     DOI: 10.1007/s10439-005-2499-y

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  24 in total

1.  Effects of stent sizing on endothelial and vessel wall stress: potential mechanisms for in-stent restenosis.

Authors:  Henry Y Chen; James Hermiller; Anjan K Sinha; Michael Sturek; Luoding Zhu; Ghassan S Kassab
Journal:  J Appl Physiol (1985)       Date:  2009-03-19

2.  Parametric Optimization of 3D Printed Hydrogel-Based Cardiovascular Stent.

Authors:  Krishna Veerubhotla; Yugyung Lee; Chi H Lee
Journal:  Pharm Res       Date:  2021-05-10       Impact factor: 4.200

3.  The stentable in vitro artery: an instrumented platform for endovascular device development and optimization.

Authors:  Elizabeth E Antoine; François P Cornat; Abdul I Barakat
Journal:  J R Soc Interface       Date:  2016-12       Impact factor: 4.118

4.  Computational fluid dynamic simulations of image-based stented coronary bifurcation models.

Authors:  Claudio Chiastra; Stefano Morlacchi; Diego Gallo; Umberto Morbiducci; Rubén Cárdenes; Ignacio Larrabide; Francesco Migliavacca
Journal:  J R Soc Interface       Date:  2013-05-15       Impact factor: 4.118

5.  The effect of resolution on viscous dissipation measured with 4D flow MRI in patients with Fontan circulation: Evaluation using computational fluid dynamics.

Authors:  Merih Cibis; Kelly Jarvis; Michael Markl; Michael Rose; Cynthia Rigsby; Alex J Barker; Jolanda J Wentzel
Journal:  J Biomech       Date:  2015-08-12       Impact factor: 2.712

6.  Bio-Adaption between Magnesium Alloy Stent and the Blood Vessel: A Review.

Authors:  Jun Ma; Nan Zhao; Lexxus Betts; Donghui Zhu
Journal:  J Mater Sci Technol       Date:  2015-12-24       Impact factor: 8.067

7.  Macro- and microscale variables regulate stent haemodynamics, fibrin deposition and thrombomodulin expression.

Authors:  Juan M Jiménez; Varesh Prasad; Michael D Yu; Christopher P Kampmeyer; Abdul-Hadi Kaakour; Pei-Jiang Wang; Sean F Maloney; Nathan Wright; Ian Johnston; Yi-Zhou Jiang; Peter F Davies
Journal:  J R Soc Interface       Date:  2014-02-19       Impact factor: 4.118

8.  Fluid-structure interaction in aortic cross-clamping: implications for vessel injury.

Authors:  Henry Y Chen; Jose A Navia; Shoaib Shafique; Ghassan S Kassab
Journal:  J Biomech       Date:  2009-11-01       Impact factor: 2.712

9.  3D fusion of intravascular ultrasound and coronary computed tomography for in-vivo wall shear stress analysis: a feasibility study.

Authors:  Alina G van der Giessen; Michiel Schaap; Frank J H Gijsen; Harald C Groen; Theo van Walsum; Nico R Mollet; Jouke Dijkstra; Frans N van de Vosse; Wiro J Niessen; Pim J de Feyter; Antonius F W van der Steen; Jolanda J Wentzel
Journal:  Int J Cardiovasc Imaging       Date:  2009-11-28       Impact factor: 2.357

10.  Luminal flow patterns dictate arterial drug deposition in stent-based delivery.

Authors:  Vijaya B Kolachalama; Abraham R Tzafriri; Davis Y Arifin; Elazer R Edelman
Journal:  J Control Release       Date:  2008-09-26       Impact factor: 9.776

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.