A prototype of a quasi-monochromatic system for mammography applications.

P Baldelli1, A Taibi, A Tuffanelli, M C Gilardoni, M Gambaccini.   

Abstract

Improvement in image contrast and dose reduction, in mammographic x-ray imaging, can be achieved using narrow energy band x-ray beams in the 16-24 keV range. As part of an Italian Government funded project, a quasi-monochromatic system for mammography applications has been developed. The system is based on a tunable narrow energy band x-ray source operating in the 16-24 keV energy range. The bremsstrahlung beam is monochromatized via Bragg diffraction by a highly oriented pyrolytic graphite mosaic crystal (HOPG). The scanning system provides a large field (18 x 24 cm2) of quasi-monochromatic x-rays with energy resolution ranging from 10% at 18 keV to 17.2% at 24 keV. The system has been characterized in terms of fluence rate and energy resolution. An x-ray tube developed ad hoc allows us to acquire images in a reasonable time to minimize the motion blur. A qualitative analysis has been performed in order to know if the prototype system performances are far from a clinical application, by evaluating the spatial resolution, the field uniformity and the image quality as a function of the quasi-monochromatic beam energy. Dose evaluation has been performed as a function of the energy and compared to a conventional system for mammography. The quasi-monochromatic prototype system can produce comparable image quality at half the dose.

Entities:  

Mesh:

Year:  2005        PMID: 15876663     DOI: 10.1088/0031-9155/50/10/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Design and characterization of a spatially distributed multibeam field emission x-ray source for stationary digital breast tomosynthesis.

Authors:  Xin Qian; Ramya Rajaram; Xiomara Calderon-Colon; Guang Yang; Tuyen Phan; David S Lalush; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

2.  Breast tomosynthesis with monochromatic beams: a feasibility study using Monte Carlo simulations.

Authors:  A Malliori; K Bliznakova; I Sechopoulos; Z Kamarianakis; B Fei; N Pallikarakis
Journal:  Phys Med Biol       Date:  2014-08-01       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.