Literature DB >> 15872133

Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB.

Koichi Yamashita1, Tomoki Nishiyama, Takeshi Yokoyama, Hidehiro Abe, Masanobu Manabe.   

Abstract

PURPOSE: To investigate the reliability of cardiac output assessed by arterial pressure waveform (PulseCO) in comparison with bolus thermodilution measurements in patients undergoing off-pump coronary artery bypass grafting (OPCAB).
METHODS: 23 patients who underwent OPCAB were enrolled in this study. After premedication with oral diazepam 10 mg, anesthesia was induced with midazolam, fentanyl and vecuronium. After induction, radial artery and pulmonary artery catheters were inserted. Cardiac output was measured simultaneously by the PulseCO and the bolus thermodilution method using the Vigilance monitor: 1) after sternotomy, 2) after opening the mediastinum, and 3) at the end of surgery. The PulseCO was calibrated initially with cardiac output determined by the thermodilution method after induction of anesthesia.
RESULTS: The correlation coefficients between the two techniques at the three measurement periods were: 1) R2 = 0.49, 2) R2 = 0.52, 3) R2 = 0.55. The limits of agreement (bias +/- 2 SD of bias) were: 1) 0.71 +/- 2.66, 2) 0.30 +/- 1.97, 3) 0.76 +/- 3.85 L.min(-1).
CONCLUSIONS: Cardiac output by PulseCO is not interchangeable with cardiac output measured by thermodilution in patients undergoing OPCAB.

Entities:  

Mesh:

Year:  2005        PMID: 15872133     DOI: 10.1007/bf03016534

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  7 in total

Review 1.  Minimally invasive monitoring of cardiac output in the cardiac surgery intensive care unit.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Giorgio della Rocca; Maxime Cannesson; Christoph K Hofer
Journal:  Curr Heart Fail Rep       Date:  2010-09

2.  Cardiac output measurement by arterial pressure waveform analysis during optimization of biventricular pacing after cardiac surgery.

Authors:  Justin H Booth; T Alexander Quinn; Marc E Richmond; Santos E Cabreriza; Alan D Weinberg; Taylor Johnston; Henry M Spotnitz
Journal:  ASAIO J       Date:  2009 Nov-Dec       Impact factor: 2.872

3.  Recent advance in patient monitoring.

Authors:  Tomoki Nishiyama
Journal:  Korean J Anesthesiol       Date:  2010-09-20

4.  Effects of vasodilation on cardiac output measured by PulseCO.

Authors:  Koichi Yamashita; Tomoki Nishiyama; Takeshi Yokoyama; Hidehiro Abe; Masanobu Manabe
Journal:  J Clin Monit Comput       Date:  2007-09-25       Impact factor: 2.502

5.  Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters.

Authors:  Mehrnaz Hadian; Hyung Kook Kim; Donald A Severyn; Michael R Pinsky
Journal:  Crit Care       Date:  2010-11-23       Impact factor: 9.097

6.  Short-term prognostic value of perioperative coronary sinus-derived-serum cardiac troponin-I, creatine kinase-MB, lactate, pyruvate, and lactate-pyruvate ratio in adult patients undergoing open heart surgery.

Authors:  Ujjwal Kumar Chowdhury; Avneesh Sheil; Poonam Malhotra Kapoor; Rajiv Narang; Parag Gharde; Vishwas Malik; Mani Kalaivani; Arindam Chaudhury
Journal:  Ann Card Anaesth       Date:  2016 Jul-Sep

7.  Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study.

Authors:  Matthias Gruenewald; Jochen Renner; Patrick Meybohm; Jan Höcker; Jens Scholz; Berthold Bein
Journal:  Crit Care       Date:  2008-10-29       Impact factor: 9.097

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.