Literature DB >> 15828300

Reoperation for a Starr-Edwards aortic prosthetic valve 33 years after initial implantation.

Takahiko Misumi1, Mikihiko Kudo, Kiyoshi Koizumi, Hiroya Kumamaru.   

Abstract

A 40-year-old woman admitted with heart failure had undergone aortic valve replacement with a Model 2310 Starr-Edwards valve due to aortic regurgitation 33 years previously. She had been followed up for several years, but discontinued follow-up and medication (including Warfarin) for the past 25 years. Echocardiography demonstrated marked dilatation and thickening of the left ventricle, and the peak pressure gradient of the prosthesis was measured as 87.9 mmHg. Under the diagnosis of chronic aortic valve prosthesis-patient mismatch with subsequent severe left ventricular dysfunction, the Starr-Edwards valve was explanted and replaced with a 23 mm St. Jude Medical prosthetic valve. The removed valve showed minimal cloth wear except for a small part of the strut. The postoperative echocardiography demonstrated recovering of left ventricular function. To our best knowledge, this case presents the longest duration for a surgically explanted Starr-Edwards aortic prosthetic valve in Japan.

Entities:  

Mesh:

Year:  2005        PMID: 15828300     DOI: 10.1007/s11748-005-0026-6

Source DB:  PubMed          Journal:  Jpn J Thorac Cardiovasc Surg        ISSN: 1344-4964


  9 in total

Review 1.  Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  J Am Coll Cardiol       Date:  2000-10       Impact factor: 24.094

2.  The natural history of adults with asymptomatic, hemodynamically significant aortic stenosis.

Authors:  P A Pellikka; R A Nishimura; K R Bailey; A J Tajik
Journal:  J Am Coll Cardiol       Date:  1990-04       Impact factor: 24.094

3.  Current status of the Starr-Edwards cloth-covered prosthetic cardiac valves.

Authors:  T Q Winter; R L Reis; D L Glancy; W C Roberts; S E Epstein; A G Morrow
Journal:  Circulation       Date:  1972-05       Impact factor: 29.690

4.  Event status of the Starr-Edwards aortic valve to 20 years: a benchmark for comparison.

Authors:  T A Orszulak; H V Schaff; F J Puga; G K Danielson; C J Mullany; B J Anderson; D M Ilstrup
Journal:  Ann Thorac Surg       Date:  1997-03       Impact factor: 4.330

5.  Thirty-year results of Starr-Edwards prostheses in the aortic and mitral position.

Authors:  O L Gödje; T Fischlein; K Adelhard; G Nollert; W Klinner; B Reichart
Journal:  Ann Thorac Surg       Date:  1997-03       Impact factor: 4.330

6.  Aortic valve replacement with Starr-Edwards valves over 14 years.

Authors:  W H Wain; P J Drury; D N Ross
Journal:  Ann Thorac Surg       Date:  1982-06       Impact factor: 4.330

7.  Forty-year survival with the Starr-Edwards heart valve prosthesis.

Authors:  Guangqiang Gao; YingXing Wu; Gary L Grunkemeier; Anthony P Furnary; Albert Starr
Journal:  J Heart Valve Dis       Date:  2004-01

8.  Konno aortoventriculoplasty in children and adolescents: from prosthetic valves to the Ross operation.

Authors:  Eldad Erez; Kirk R Kanter; Vincent K H Tam; Willis H Williams
Journal:  Ann Thorac Surg       Date:  2002-07       Impact factor: 4.330

9.  Up to thirty-year survival after aortic valve replacement in the small aortic root.

Authors:  G W He; G L Grunkemeier; H L Gately; A P Furnary; A Starr
Journal:  Ann Thorac Surg       Date:  1995-05       Impact factor: 4.330

  9 in total
  1 in total

1.  Use of the Bentall method for replacement of an aortic Starr-Edwards ball valve prosthesis implanted 32 years previously.

Authors:  Daisuke Yotsumoto; Yoshifumi Iguro; Takayuki Ueno; Kazuhisa Matsumoto; Ryuzo Sakata
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-08
  1 in total