Current-level discrimination using bipolar and monopolar electrode configurations in cochlear implants.

Ward R Drennan1, Bryan E Pfingst.   

Abstract

This study examined current-level discrimination ability in listeners with cochlear implants using bipolar and monopolar electrode configurations. Current-level discrimination ability was measured as a function of electrode configuration (monopolar and bipolar), stimulation site (8 and 16) and level (5%, 15%, 25%, 50% and 80% of the dynamic range). Weber fractions usually decreased with increasing level. Differences in Weber fractions between monopolar and bipolar configurations were observed for listeners with short durations of deafness (<5 years). For these listeners, in the bipolar condition at the more-apical site 16, Weber fractions remained constant with increasing level, and the Weber fractions at low levels were smaller than in other conditions. We suggest that nerve density was better and the nerve-to-site-of-action-potential distance was smaller in these cases such that more fibers could be recruited with a unit increase in current level, leading to better current-level sensitivity.

Entities:  

Mesh:

Year:  2005        PMID: 15811709     DOI: 10.1016/j.heares.2004.11.010

Source DB:  PubMed          Journal:  Hear Res        ISSN: 0378-5955            Impact factor:   3.208


  7 in total

1.  Current-level discrimination in the context of interleaved, multichannel stimulation in cochlear implants: effects of number of stimulated electrodes, pulse rate, and electrode separation.

Authors:  Ward R Drennan; Bryan E Pfingst
Journal:  J Assoc Res Otolaryngol       Date:  2006-06-21

2.  Spatially distinct functional output regions within the central nucleus of the inferior colliculus: implications for an auditory midbrain implant.

Authors:  Hubert H Lim; David J Anderson
Journal:  J Neurosci       Date:  2007-08-08       Impact factor: 6.167

3.  Intensity coding in electric hearing: effects of electrode configurations and stimulation waveforms.

Authors:  Tiffany Elise H Chua; Mark Bachman; Fan-Gang Zeng
Journal:  Ear Hear       Date:  2011 Nov-Dec       Impact factor: 3.570

4.  Comparison of Intensity Discrimination between Children Using Cochlear Implants and Typically Developing Children.

Authors:  Shubha Tak; Asha Yathiraj
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

5.  Current-level discrimination and spectral profile analysis in multi-channel electrical stimulation.

Authors:  Matthew J Goupell; Bernhard Laback; Piotr Majdak; Wolf-Dieter Baumgartner
Journal:  J Acoust Soc Am       Date:  2008-11       Impact factor: 1.840

6.  Comparisons between detection threshold and loudness perception for individual cochlear implant channels.

Authors:  Julie Arenberg Bierer; Amberly D Nye
Journal:  Ear Hear       Date:  2014 Nov-Dec       Impact factor: 3.570

7.  Extending the limits of place and temporal pitch perception in cochlear implant users.

Authors:  Olivier Macherey; John M Deeks; Robert P Carlyon
Journal:  J Assoc Res Otolaryngol       Date:  2010-11-30