On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies.

Thomas Beyer1, Lutz Tellmann, Ingo Nickel, Uwe Pietrzyk.   

Abstract

UNLABELLED: Involuntary patient motion from insufficient patient preparation may lead to local misregistration of PET/CT images and, thus, can invalidate the attempt to fuse the resulting images. We estimate the efficacy of selected patient support structures in reducing the likelihood of patient motion in the area of the head and neck during whole-body PET/CT studies.
METHODS: Motion of the head and neck was estimated in 51 healthy volunteers during simulated whole-body PET/CT studies using an infrared camera-based tracking system. Four patient positioning schemes (arms down) were studied, with the neck placed on a standard PET head holder with no support at the sides (setup A), on a special head holder fitted with a subject-specific mold from construction foam (setup B), on a vacuum-lock bag (setup C), and on a special head holder fitted with a vacuum-lock bag (setup D). We report the average motion of the head and neck as the difference in the position of a set of target points between the simulated CT image and PET image of the head and neck. To estimate the efficacy of additional patient support measures in clinical practice, we reviewed the misregistration of the head and neck in whole-body PET/CT studies of 10 patients each who were imaged using setups A and C by comparing the mean translational and rotational alignment parameters from a semiautomatic linear registration approach needed to realign the CT and PET images.
RESULTS: Average translational and rotational misalignment of the head and neck was highest for setup A, at 7 mm and 1 degrees , respectively. Misalignment was reduced to a minimum of 1.4 mm and 0.3 degrees for setup D. Setup B resulted in a similar reduction in patient motion of the head and neck: 2.4 mm and 0.4 degrees , whereas setup C provided only somewhat improved support, with a resulting average misalignment of 4.5 mm and 0.7 degrees. In clinical PET/CT, we found setup C to reduce translational misalignment of the CT and PET images of the head and neck to 2 mm, compared with 6 mm for setup A, whereas no significant reduction of rotational misalignment was observed.
CONCLUSION: Average motion of the head and neck in unrestrained subjects during whole-body PET/CT examinations can be reduced by use of rigid positioning aids, such as foam molds, or vacuum-lock bags. Vacuum-lock bags are reusable, quickly adaptable, and olfactory neutral and can be used routinely, either alone or in combination with a head holder, in whole-body PET/CT for high-quality examinations.

Entities:  

Mesh:

Year:  2005        PMID: 15809481

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  13 in total

1.  Respiration artifacts in whole-body (18)F-FDG PET/CT studies with combined PET/CT tomographs employing spiral CT technology with 1 to 16 detector rows.

Authors:  Thomas Beyer; Sandra Rosenbaum; Patrick Veit; Jörg Stattaus; Stefan P Müller; Frank P Difilippo; Heiko Schöder; Osama Mawlawi; Fiona Roberts; Andreas Bockisch; Hilmar Kühl
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-08-26       Impact factor: 9.236

2.  The effect of head movement on CT perfusion summary maps: simulations with CT hybrid phantom data.

Authors:  F Fahmi; A Riordan; L F M Beenen; G J Streekstra; N Y Janssen; H W de Jong; C B L Majoie; E van Bavel; H A Marquering
Journal:  Med Biol Eng Comput       Date:  2013-10-30       Impact factor: 2.602

3.  3D movement correction of CT brain perfusion image data of patients with acute ischemic stroke.

Authors:  Fahmi Fahmi; Henk A Marquering; Jordi Borst; Geert J Streekstra; Ludo F M Beenen; Joris M Niesten; Birgitta K Velthuis; Charles B L Majoie; Ed vanBavel
Journal:  Neuroradiology       Date:  2014-04-09       Impact factor: 2.804

4.  SPECT/CT workflow and imaging protocols.

Authors:  Catherine Beckers; Roland Hustinx
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-12-07       Impact factor: 9.236

5.  The effect of patient positioning aids on PET quantification in PET/MR imaging.

Authors:  Frederic Mantlik; Matthias Hofmann; Matthias K Werner; Alexander Sauter; Jürgen Kupferschläger; Bernhard Schölkopf; Bernd J Pichler; Thomas Beyer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-02-10       Impact factor: 9.236

6.  TNM staging with FDG-PET/CT in patients with primary head and neck cancer.

Authors:  Patrick Veit-Haibach; Christopher Luczak; Isabel Wanke; Markus Fischer; Thomas Egelhof; Thomas Beyer; Gerlinde Dahmen; Andreas Bockisch; Sandra Rosenbaum; Gerald Antoch
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-08-24       Impact factor: 9.236

7.  An automated algorithm to identify and quantify brown adipose tissue in human 18F-FDG-PET/CT scans.

Authors:  Megan R Ruth; Tyler Wellman; Gustavo Mercier; Thomas Szabo; Caroline M Apovian
Journal:  Obesity (Silver Spring)       Date:  2013-08       Impact factor: 5.002

8.  Off-line motion correction methods for multi-frame PET data.

Authors:  Jurgen E M Mourik; Mark Lubberink; Floris H P van Velden; Adriaan A Lammertsma; Ronald Boellaard
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-12       Impact factor: 9.236

9.  Nonrigid image registration for head and neck cancer radiotherapy treatment planning with PET/CT.

Authors:  Rob H Ireland; Karen E Dyker; David C Barber; Steven M Wood; Michael B Hanney; Wendy B Tindale; Neil Woodhouse; Nigel Hoggard; John Conway; Martin H Robinson
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-04-18       Impact factor: 7.038

10.  Physiologic [18F]fluorodeoxyglucose uptake of floor of mouth muscles in PET/CT imaging: a problem of body position during FDG uptake?

Authors:  Stephan K Haerle; Thomas F Hany; Nader Ahmad; Irene Burger; Gerhard F Huber; Daniel T Schmid
Journal:  Cancer Imaging       Date:  2013-02-20       Impact factor: 3.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.