Literature DB >> 15803068

Clinical accuracy of fluoroscopic computer-assisted pedicle screw fixation: a CT analysis.

Y Raja Rampersaud1, Justin H T Pik, David Salonen, Samina Farooq.   

Abstract

STUDY
DESIGN: Observational cohort study with computerized tomography (CT) analysis of in vivo pedicle screw placement.
OBJECTIVE: To evaluate the clinical accuracy of computer-assisted fluoroscopy for the placement of thoracic and lumbosacral (LS) pedicle screws. SUMMARY OF BACKGROUND DATA: Computer-assisted fluoroscopy is an adjunct for the placement of pedicle screws.
METHODS: Postoperative CT of 360 titanium pedicle screws (281 LS [L1-S1]; 79 thoracic [T2-T12]) were independently assessed. All screws were placed using the FluoroNav system (Medtronic Surgical Navigation Technologies, Louisville, CO). The relative position of the screw to the pedicle was graded as follows: A = completely in; B = <2 mm breach; C = 2-4 mm breach; and D = >4 mm breach. If an osseous breach occurred, the direction of the breach was further classified.
RESULTS: Eight-five percent of screws were completely contained within the pedicle. Fifty-five-pedicle breaches (25 medial; 30 lateral) occurred. Pedicle breaches were graded B in 13.1% (47 breaches), C in 1.9% (7), and D in 0.3% (1) of screws. Pedicle breach was significantly higher in the thoracic compared to the LS spine, 31.6% (25/79) and 10.6% (30/281), respectively (P < 0.0001). In the thoracic spine, 72% of pedicle breaches were lateral. In 49% of breaches, the screw diameter was larger than the pedicle diameter. None of the pedicle breaches were associated with neurologic or other clinical sequelae.
CONCLUSIONS: The clinical pedicle breach rate in this study is comparable to those reported using conventional techniques with or without fluoroscopic assistance. FluoroNav appears to be a safe adjunct for the placement of thoracic and LS pedicle screws.

Mesh:

Year:  2005        PMID: 15803068     DOI: 10.1097/01.brs.0000157490.65706.38

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  40 in total

1.  Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: a cadaveric study.

Authors:  Claudio Lamartina; Riccardo Cecchinato; Zsolt Fekete; Alberto Lipari; Meinrad Fiechter; P Berjano
Journal:  Eur Spine J       Date:  2015-10-23       Impact factor: 3.134

2.  O-arm(®)-based spinal navigation and intraoperative 3D-imaging: first experiences.

Authors:  O Gonschorek; S Hauck; U Spiegl; T Weiß; R Pätzold; V Bühren
Journal:  Eur J Trauma Emerg Surg       Date:  2011-03-31       Impact factor: 3.693

3.  3D-based navigation in posterior stabilisations of the cervical and thoracic spine: problems and benefits. Results of 451 screws.

Authors:  J-S Jarvers; S Katscher; A Franck; S Glasmacher; C Schmidt; T Blattert; C Josten
Journal:  Eur J Trauma Emerg Surg       Date:  2011-04-01       Impact factor: 3.693

4.  Observer reliability in evaluating pedicle screw placement using computed tomography.

Authors:  Victor Kosmopoulos; Nicolas Theumann; Stefano Binaghi; Constantin Schizas
Journal:  Int Orthop       Date:  2006-09-12       Impact factor: 3.075

5.  Electrical conductivity measurement: a new technique to detect iatrogenic initial pedicle perforation.

Authors:  Ciaran Bolger; Michael O Kelleher; Linda McEvoy; M Brayda-Bruno; A Kaelin; J-Y Lazennec; J-C Le Huec; C Logroscino; P Mata; P Moreta; G Saillant; R Zeller
Journal:  Eur Spine J       Date:  2007-06-30       Impact factor: 3.134

6.  Image-guided pedicle screw insertion accuracy: a meta-analysis.

Authors:  Nai-Feng Tian; Hua-Zi Xu
Journal:  Int Orthop       Date:  2009-05-08       Impact factor: 3.075

Review 7.  Robot-assisted and fluoroscopy-guided pedicle screw placement: a systematic review.

Authors:  Hani J Marcus; Thomas P Cundy; Dipankar Nandi; Guang-Zhong Yang; Ara Darzi
Journal:  Eur Spine J       Date:  2013-06-26       Impact factor: 3.134

Review 8.  Functional outcome of computer-assisted spinal pedicle screw placement: a systematic review and meta-analysis of 23 studies including 5,992 pedicle screws.

Authors:  Rajeev Verma; Sonal Krishan; Kurt Haendlmayer; A Mohsen
Journal:  Eur Spine J       Date:  2010-01-06       Impact factor: 3.134

Review 9.  Image-guided spine surgery: state of the art and future directions.

Authors:  Thorsten Tjardes; Sven Shafizadeh; Dieter Rixen; Thomas Paffrath; Bertil Bouillon; Eva S Steinhausen; Holger Baethis
Journal:  Eur Spine J       Date:  2009-09-11       Impact factor: 3.134

10.  Technical Report of Free Hand Pedicle Screw Placement using the Entry Points with Junction of Proximal Edge of Transverse Process and Lamina in Lumbar Spine: Analysis of 2601 Consecutive Screws.

Authors:  Chang Hyun Oh; Seung Hwan Yoon; Yongjung J Kim; Dongkeun Hyun; Hyeong-Chun Park
Journal:  Korean J Spine       Date:  2013-03-31