Accuracy of segmentation of a commercial computer-aided detection system for mammography.

Jay A Baker1, Eric L Rosen, Michele M Crockett, Joseph Y Lo.   

Abstract

PURPOSE: To assess the accuracy of segmentation in a commercially available computer-aided detection (CAD) system.
MATERIALS AND METHODS: Approval for this study was obtained from the authors' institutional review board. Informed consent was not required by the board for this review, as data were stripped of patient identifiers. Two thousand twenty mammograms from 507 women were analyzed with the hardware and software of a commercial CAD system. The accuracy of the segmentation process was determined semiquantitatively and categorized as near perfect if the skin line of the breast was accurately detected, acceptable if only subcutaneous fat was excluded, or unacceptable if any breast parenchyma was excluded from consideration. The accuracy of segmentation was compared for different breast densities and film sizes by using logistic regression (P < .05).
RESULTS: Overall, segmentation was near perfect or acceptable in almost 96.8% of images. However, segmentation defects were significantly more common in mammograms with heterogeneously dense breast tissue (8% unacceptable) than in those with fatty replaced (0% unacceptable), scattered (1.2% unacceptable), or extremely dense (1.8% unacceptable) breast parenchyma (P < .05). For images with unacceptable segmentation, the average percentage of breast parenchyma excluded was almost 25% (range, 5%-100%), with no significant differences among breast densities.
CONCLUSION: For one commercial CAD system, segmentation was usually near perfect or acceptable but was unacceptable more than five times more frequently for mammograms of breasts with heterogeneously dense parenchyma than for those with all other breast densities. On average, one-quarter of the breast parenchyma was excluded from CAD analysis for images with unacceptable segmentation. (c) RSNA, 2005.

Entities:  

Mesh:

Year:  2005        PMID: 15770043     DOI: 10.1148/radiol.2352040899

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

Review 1.  Review of recent advances in segmentation of the breast boundary and the pectoral muscle in mammograms.

Authors:  Mario Mustra; Mislav Grgic; Rangaraj M Rangayyan
Journal:  Med Biol Eng Comput       Date:  2015-11-06       Impact factor: 2.602

2.  ACR-AAPM-SIIM practice guideline for determinants of image quality in digital mammography.

Authors:  Kalpana M Kanal; Elizabeth Krupinski; Eric A Berns; William R Geiser; Andrew Karellas; Martha B Mainiero; Melissa C Martin; Samir B Patel; Daniel L Rubin; Jon D Shepard; Eliot L Siegel; Judith A Wolfman; Tariq A Mian; Mary C Mahoney
Journal:  J Digit Imaging       Date:  2013-02       Impact factor: 4.056

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.