Literature DB >> 15746858

In vitro assessment of image-guided otologic surgery: submillimeter accuracy within the region of the temporal bone.

Robert F Labadie1, Rohan J Shah, Steve S Harris, Ebru Cetinkaya, David S Haynes, Michael R Fenlon, Andrzej S Juszczyk, Robert L Galloway, J Michael Fitzpatrick.   

Abstract

OBJECTIVES: Application of image-guided surgery to otology has been limited by the need for submillimeter accuracy via a fiducial system that is easily usable (noninvasive and nonobstructive).
METHODS: A dental bite-block was fitted with a rigid frame with 7 fiducial markers surrounding each external ear. The temporal bones of 3 cadaveric skulls were removed and replaced with surgical targets arranged in a bull's-eye pattern about the centroid of each temporal bone. The surgical targets were identified both within CT scans and in physical space using an infrared optical tracking system. The difference between positions in CT space versus physical space was calculated as target registration error.
RESULTS: A total of 234 independent target registration errors were calculated. Mean +/- standard deviation = 0.73 mm +/- 0.25 mm.
CONCLUSIONS: These findings show that image-guided otologic surgery with submillimeter accuracy is achievable with a minimally invasive fiducial frame. Significance In vivo validation of the system is ongoing. With such validation, this system may facilitate clinically applicable image-guided otologic surgery. EBM RATING: A.

Mesh:

Year:  2005        PMID: 15746858     DOI: 10.1016/j.otohns.2004.09.141

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  14 in total

1.  Determination of the curling behavior of a preformed cochlear implant electrode array.

Authors:  Thomas S Rau; Omid Majdani; Andreas Hussong; Thomas Lenarz; Martin Leinung
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-07-28       Impact factor: 2.924

2.  Is there an equivalence of non-invasive to invasive referenciation in computer-aided surgery?

Authors:  Tanja D Grauvogel; Juergen Grauvogel; Susan Arndt; Ansgar Berlis; Wolfgang Maier
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-05-05       Impact factor: 2.503

3.  Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model.

Authors:  Thomas S Rau; Andreas Hussong; Martin Leinung; Thomas Lenarz; Omid Majdani
Journal:  Int J Comput Assist Radiol Surg       Date:  2009-04-10       Impact factor: 2.924

4.  Analysis of surgeon's line of sight using an optical tracking system with a multifaceted marker device.

Authors:  Hiroaki Tsunezuka; Daishiro Kato; Kunihiko Terauchi; Masanori Shimomura; Kaori Ichise; Kazuhiro Ito; Atsushi Nishikawa; Junichi Shimada
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-04-22       Impact factor: 2.924

5.  Warning navigation system using real-time safe region monitoring for otologic surgery.

Authors:  Byunghyun Cho; Masamichi Oka; Nozomu Matsumoto; Riichi Ouchida; Jaesung Hong; Makoto Hashizume
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-11-16       Impact factor: 2.924

6.  A projected landmark method for reduction of registration error in image-guided surgery systems.

Authors:  Nasim Dadashi Serej; Alireza Ahmadian; Saeed Mohagheghi; Seyed Musa Sadrehosseini
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-05-28       Impact factor: 2.924

7.  Automated dental implantation using image-guided robotics: registration results.

Authors:  Xiaoyan Sun; Frederic D McKenzie; Sebastian Bawab; Jiang Li; Yongki Yoon; Jen-K Huang
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-08       Impact factor: 2.924

8.  SOX2 is a dose-dependent regulator of retinal neural progenitor competence.

Authors:  Olena V Taranova; Scott T Magness; B Matthew Fagan; Yongqin Wu; Natalie Surzenko; Scott R Hutton; Larysa H Pevny
Journal:  Genes Dev       Date:  2006-05-01       Impact factor: 11.361

9.  Accuracy of image-guided surgical systems at the lateral skull base as clinically assessed using bone-anchored hearing aid posts as surgical targets.

Authors:  Ramya Balachandran; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otol Neurotol       Date:  2008-12       Impact factor: 2.311

10.  Quantitative error analysis for computer assisted navigation: a feasibility study.

Authors:  Ö Güler; M Perwög; F Kral; F Schwarm; Z R Bárdosi; G Göbel; W Freysinger
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.