Lu Wang1, Rojymon Jacob, Lili Chen, Charlie Ma, Benjamin Movsas, Steve Feigenberg, Andre Konski.
Abstract
The purpose of this work is to prospectively assess the setup accuracy that can be achieved with a stereotactic body localizer (SBL) in immobilizing patients for stereotactic intensity-modulated radiotherapy (IMRT) for prostate cancer. By quantifying this important factor and target mobility in the SBL, we expect to provide a guideline for selecting planning target volume margins for stereotactic treatment planning. We analyzed data from 40 computed tomography (CT) studies (with slice thickness of 3 mm) involving 10 patients with prostate cancer. Each patient had four sets of CT scans during the course of radiotherapy. For the purpose of this study, all four sets of CT scans were obtained with the patients immobilized in a customized body pillow formed by vacuum suction. Unlike other immobilization devices, this system consists not only of a customized body pillow, but also of a fixation sheet used to suppress patient respiratory motion, a stereotactic body frame to provide stereotaxy, and a carbon fiber base board to which both the body cushion and the frame are affixed. We identified four bony landmarks and measured their coordinates in the stereotactic body frame on each set of CT scans. The displacements of the bony landmarks from their corresponding positions on the simulation scan (first CT scan) were analyzed in three dimensions in terms of overall, systematic, and random categories. The initial planned isocenter was also marked on the patients' skin with fiducials for each CT study. The distance from each bony landmark to the fiducial-based isocenter was measured and compared among the four sets of CT scans. The deviations in distances were also compared to those measured from the landmarks to the stereotactic frame center, in order to determine the effectiveness of the rigid body frame in positioning patients with prostate cancer. Target inter-fraction motion in this system was also studied for five patients by measuring the deviations in distances from the target geometric center to the bony landmarks. Our results showed that the overall setup accuracy had standard deviations (SDs) of 2.58 mm, 2.41 mm, and 3.51 mm in lateral (LAT), anterior-posterior(AP), and superior-inferior (SI) directions, respectively. The random component had SDs of 1.72 mm, 2.06 mm, and 2.79 mm, and the systematic component showed SDs of 0.92 mm, -0.27 mm, and 0.90 mm in these three directions. In terms of three-dimensional vector, the mean displacement over 116 measurements was 3.0 mm with an SD of 1.29 mm. Compared to the rigid reference, the skin-mark-based reference was less reliable for patient repositioning in terms of reproducing known bony landmark positions. The mean target mobility relative to the bony landmarks was 2.22 +/- 3.45 mm, 0.17 +/- 1.11 mm, and 0.11 +/- 2.69 mm in the AP, LAT, and SI directions, respectively. In conclusion, the body immobilization system has the ability to immobilize prostate cancer patients with satisfactory setup accuracy for fractionated extracranial stereotactic radiotherapy. A rigid frame system serves as a reliable alignment reference in terms of repositioning patients into the planning position, while skin-based reference showed larger deviations in repositioning patients.Entities:
Mesh:
Year: 2004 PMID: 15738910 PMCID: PMC5723461 DOI: 10.1120/jacmp.v5i2.1947
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The components of the body localizer system. (a) The base board, body frame, and body pillow; (b) patient immobilized in the body localizer.
Figure 2The body localizer consisting of nine rods shown on one slice of the CT image and the coordinate system established by the rods.
Figure 3Systematic errors and the range of random errors for individual patients.
Average displacement and standard deviation (SD) of the overall distribution and the distribution of systematic and random displacements along the three main axes for the whole group
| Distribution of displacements (1 SD (mm)) | ||||
|---|---|---|---|---|
| Direction | Average displacement (mm) ( | Overall ( | Systematic ( | Random ( |
| lateral | 0.96 | 2.58 | 1.95 | 1.72 |
| anterior‐posterior | –0.22 | 2.41 | 1.20 | 2.06 |
| superior‐inferior | 0.86 | 3.51 | 2.12 | 2.79 |
Figure 4Frequency and cumulative probability of the displacements along the LAT (x), AP (y), and SI (z) directions for total of 116 measurements.
Figure 5Cumulative distribution of three‐dimensional vector for overall displacement, calculated from 116 sets of displacements for 10 patients.
Comparisons of patient positioning accuracy determined with respect to two different references (the SBL and the BB‐based) in terms of average displacements and the distribution of systematic errors in the three major directions
| Average displacement (mm) ( | 1 SD for systematic error (mm) ( | Maximum displacements (mm) | ||||
|---|---|---|---|---|---|---|
| Directions | Relative to BL | Relative to BB | Relative to BL | Relative to BB | Relative to BL | Relative to BB |
| left‐right | 0.92 | –1.41 | 1.95 | 3.47 | 6.8 | 15.0 |
| anterior‐posterior | –0.27 | –0.42 | 1.20 | 5.11 | 7.3 | 13.55 |
| superior‐inferior | 0.90 | –0.83 | 2.12 | 3.15 | 9.9 | 11.3 |
Figure 6Frequencies and cumulative distributions of the target motion in respect to the patient skeleton along AP, LAT, and SI directions.