Three dimensional variability in patient positioning using bite block immobilization in 3D-conformal radiation treatment for ENT-tumors.

J Willner1, U Hädinger, M Neumann, F J Schwab, K Bratengeier, M Flentje.   

Abstract

BACKGROUND AND
PURPOSE: The aim of this prospective study was to analyze the three-dimensional (3D) reproducibility of the isocenter position and of patient positioning with the use of bite block immobilization by means of a simple verification procedure for a complex beam arrangement applied for ENT-tumors.
MATERIALS AND METHODS: We analyzed the positioning data of 29 consecutive patients treated for ENT-tumors at the Department of Radiotherapy and Oncology of the University of Wurzburg. A total of 136 treatment sessions were analyzed. Patients were positioned and immobilized using an individualized bite block system and a head and neck support. A complex beam arrangement was applied combining two offset rotational and two oblique wedge fields on a 5 MV linear accelerator. Orthogonal verification films were taken once weekly. Four to six film pairs per patient were obtained (during 4-6 weeks) with a mean number of 4.7 film pairs per patient. These were compared to the corresponding orthogonal simulator films taken during primary simulation. Deviations of the verified isocenter from the isocenter on the simulator film were measured and analyzed in three dimensions in terms of overall, systematic and random categories. A 3D-deviation vector was calculated from these 3D data as well as a 2D-deviation vector (for comparison with literature data) from the lateral verification films.
RESULTS: The overall setup deviation showed standard deviations (SD) of 2.5, 2.7 and 3.1 mm along the cranio-caudal, anterior-posterior and medio-lateral axes, respectively. The random component ranged from SD 1.9 to 2.1 mm and the systematic component ranged from SD 1.8 to 2.2 mm. The mean length of the 3D-vector was 3.1 mm for the systematic as well as the random component. Ninety percent of 3D systematic and random deviations were less than 5 mm. The mean length of the 2D-vector was 2.4 mm for the random component and 2.2 mm for the systematic component. Ninety percent of 2D-random and systematic variations were less than 4 mm.
CONCLUSIONS: The presented individualized bite block immobilization device provides an accurate and reproducible patient positioning for 3D-conformal radiation therapy in the head and neck. Random and systematic deviations in each of the three directions are in the range of +/-4 mm (2 SD, comprising 95% of the deviations) and are within the range or even less than deviations described for most thermoplastic or PVC-mask fixation devices. These deviations should be taken into account during definition of planning target volume in head and neck tumors.

Entities:  

Mesh:

Year:  1997        PMID: 9215794     DOI: 10.1016/s0167-8140(97)00055-8

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  8 in total

Review 1.  Current progress in adaptive radiation therapy for head and neck cancer.

Authors:  David L Schwartz
Journal:  Curr Oncol Rep       Date:  2012-04       Impact factor: 5.075

2.  Variable planning margin approach to account for locoregional variations in setup uncertainties.

Authors:  Jinzhong Yang; Adam S Garden; Yongbin Zhang; Lifei Zhang; Lei Dong
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

3.  A newly developed patient fixation system using a dedicated mouthpiece and dental impression materials for head and neck radiotherapy: a preliminary study.

Authors:  Masataka Sakamoto; Kenta Konishi; Keiichi Ohira; Masanori Hirata; Kohei Wakabayashi; Shuhei Aramaki; Ryo Kokubo; Katsumasa Nakamura
Journal:  J Radiat Res       Date:  2022-09-21       Impact factor: 2.438

4.  Utility of intraoral stents in external beam radiotherapy for head and neck cancer.

Authors:  Hiroshi Doi; Masao Tanooka; Toshihisa Ishida; Kuniyasu Moridera; Kenji Ichimiya; Kazuo Tarutani; Kazuhiro Kitajima; Masayuki Fujiwara; Hiromitsu Kishimoto; Norihiko Kamikonya
Journal:  Rep Pract Oncol Radiother       Date:  2017-05-08

5.  Adaptive radiation therapy for head and neck cancer-can an old goal evolve into a new standard?

Authors:  David L Schwartz; Lei Dong
Journal:  J Oncol       Date:  2010-08-18       Impact factor: 4.375

6.  Stereotactic IMRT for prostate cancer: setup accuracy of a new stereotactic body localization system.

Authors:  Lu Wang; Rojymon Jacob; Lili Chen; Charlie Ma; Benjamin Movsas; Steve Feigenberg; Andre Konski
Journal:  J Appl Clin Med Phys       Date:  2004-04-01       Impact factor: 2.102

7.  Accuracy and Precision of a Veterinary Neuronavigation System for Radiation Oncology Positioning.

Authors:  Isabelle F Vanhaezebrouck; Elizabeth A Ballegeer; Stephen Frey; Rob Sieffert
Journal:  J Vet Med       Date:  2018-02-13

8.  Recent advances in image-guided radiotherapy for head and neck carcinoma.

Authors:  Sameer K Nath; Daniel R Simpson; Brent S Rose; Ajay P Sandhu
Journal:  J Oncol       Date:  2009-07-29       Impact factor: 4.375

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.