Literature DB >> 15736247

Invasive human magnetic resonance imaging: feasibility during revascularization in a combined XMR suite.

Alexander J Dick1, Venkatesh K Raman, Amish N Raval, Michael A Guttman, Richard B Thompson, Cengizhan Ozturk, Dana C Peters, Annette M Stine, Victor J Wright, William H Schenke, Robert J Lederman.   

Abstract

We tested the feasibility and safety of invasive magnetic resonance imaging (MRI) during peripheral angioplasty. Real-time MRI can image soft tissue and may potentially guide therapeutic procedures without ionizing radiation or nephrotoxic contrast. MRI-guided diagnostic catheterization has been described recently, but safe and conspicuous catheter devices are not widely available. An active guidewire, which serves as an MRI receiver antenna, might be useful to guide catheterization or even to image atheroma. We describe a combined interventional suite offering both X-ray fluoroscopy and real-time MRI. We used a 0.030'' active guidewire receiver coil for invasive MRI after X-ray lesion traversal in patients undergoing percutaneous iliofemoral artery revascularization. Intravascular MRI was compared with noninvasive MRI, X-ray angiography, and intravascular ultrasound (IVUS). Seven eligible patients consented to participate, but three were excluded because of lengthy revascularization procedures. Four remaining patients safely underwent combined X-ray fluoroscopy and real-time magnetic resonance imaging (XMR) transport, continuous monitoring, and all imaging modalities. There was no device dislodgment, contamination or evidence of heating. The intravascular MRI coil was well visualized except at the tip, but did not provide superior mural imaging compared with IVUS. Therefore, because an adequate safety and workflow experience was obtained, enrollment was terminated after only four subjects. Invasive MRI is feasible and apparently safe during peripheral angioplasty. Patients can safely be transported and monitored in an XMR interventional suite. An active quarter-wavelength guidewire coil does not provide superior imaging compared with IVUS, but provides satisfactory guidewire visualization. These tools may prove useful for advanced therapeutic procedures in the future.

Entities:  

Mesh:

Year:  2005        PMID: 15736247      PMCID: PMC1317610          DOI: 10.1002/ccd.20302

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  23 in total

1.  Real-time volume rendered MRI for interventional guidance.

Authors:  Michael A Guttman; Robert J Lederman; Jonathan M Sorger; Elliot R McVeigh
Journal:  J Cardiovasc Magn Reson       Date:  2002       Impact factor: 5.364

2.  Active device tracking and high-resolution intravascular MRI using a novel catheter-based, opposed-solenoid phased array coil.

Authors:  Claudia M Hillenbrand; Daniel R Elgort; Eddy Y Wong; Arne Reykowski; Frank K Wacker; Jonathan S Lewin; Jeffrey L Duerk
Journal:  Magn Reson Med       Date:  2004-04       Impact factor: 4.668

3.  Cardiovascular catheters and accessories: ex vivo testing of ferromagnetism, heating, and artifacts associated with MRI.

Authors:  F G Shellock; V J Shellock
Journal:  J Magn Reson Imaging       Date:  1998 Nov-Dec       Impact factor: 4.813

4.  An expandable intravenous RF coil for arterial wall imaging.

Authors:  A J Martin; R F McLoughlin; K C Chu; E A Barberi; B K Rutt
Journal:  J Magn Reson Imaging       Date:  1998 Jan-Feb       Impact factor: 4.813

5.  Active MR visualization of a vascular guidewire in vivo.

Authors:  M E Ladd; G G Zimmermann; H H Quick; J F Debatin; P Boesiger; G K von Schulthess; G C McKinnon
Journal:  J Magn Reson Imaging       Date:  1998 Jan-Feb       Impact factor: 4.813

6.  Abdominal MR imaging with a volumetric interpolated breath-hold examination.

Authors:  N M Rofsky; V S Lee; G Laub; M A Pollack; G A Krinsky; D Thomasson; M M Ambrosino; J C Weinreb
Journal:  Radiology       Date:  1999-09       Impact factor: 11.105

7.  In vivo intravascular MR imaging: transvenous technique for arterial wall imaging.

Authors:  Lawrence V Hofmann; Robert P Liddell; Aravind Arepally; Brian Montague; Xiaoming Yang; David A Bluemke
Journal:  J Vasc Interv Radiol       Date:  2003-10       Impact factor: 3.464

8.  Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in Swine.

Authors:  Alexander J Dick; Michael A Guttman; Venkatesh K Raman; Dana C Peters; Breno S S Pessanha; Jonathan M Hill; Scott Smith; Greig Scott; Elliot R McVeigh; Robert J Lederman
Journal:  Circulation       Date:  2003-12-01       Impact factor: 29.690

9.  Cardiac catheterisation guided by MRI in children and adults with congenital heart disease.

Authors:  Reza Razavi; Derek L G Hill; Stephen F Keevil; Marc E Miquel; Vivek Muthurangu; Sanjeet Hegde; Kawal Rhode; Michael Barnett; Joop van Vaals; David J Hawkes; Edward Baker
Journal:  Lancet       Date:  2003-12-06       Impact factor: 79.321

Review 10.  Endovascular interventional magnetic resonance imaging.

Authors:  L W Bartels; C J G Bakker
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

View more
  19 in total

Review 1.  CMR for characterization of the myocardium in acute coronary syndromes.

Authors:  Erica Dall'Armellina; Theodoros D Karamitsos; Stefan Neubauer; Robin P Choudhury
Journal:  Nat Rev Cardiol       Date:  2010-09-21       Impact factor: 32.419

Review 2.  Cardiovascular interventional magnetic resonance imaging.

Authors:  Robert J Lederman
Journal:  Circulation       Date:  2005-11-08       Impact factor: 29.690

3.  Real-time magnetic resonance imaging to guide pediatric endovascular procedures.

Authors:  A N Raval; R J Lederman
Journal:  Pediatr Cardiol       Date:  2005 May-Jun       Impact factor: 1.655

Review 4.  Advances in interventional cardiovascular MRI.

Authors:  Venkatesh K Raman; Robert J Lederman
Journal:  Curr Cardiol Rep       Date:  2006-02       Impact factor: 2.931

5.  Interventional cardiovascular procedures guided by real-time MR imaging: an interactive interface using multiple slices, adaptive projection modes and live 3D renderings.

Authors:  Michael A Guttman; Cenghizhan Ozturk; Amish N Raval; Venkatesh K Raman; Alexander J Dick; Ranil DeSilva; Parag Karmarkar; Robert J Lederman; Elliot R McVeigh
Journal:  J Magn Reson Imaging       Date:  2007-12       Impact factor: 4.813

6.  Real-time MRI guided atrial septal puncture and balloon septostomy in swine.

Authors:  Amish N Raval; Parag V Karmarkar; Michael A Guttman; Cengizhan Ozturk; Ranil Desilva; Ronnier J Aviles; Victor J Wright; William H Schenke; Ergin Atalar; Elliot R McVeigh; Robert J Lederman
Journal:  Catheter Cardiovasc Interv       Date:  2006-04       Impact factor: 2.692

7.  Real-time magnetic resonance imaging-guided endovascular recanalization of chronic total arterial occlusion in a swine model.

Authors:  Amish N Raval; Parag V Karmarkar; Michael A Guttman; Cengizhan Ozturk; Smita Sampath; Ranil DeSilva; Ronnier J Aviles; Minnan Xu; Victor J Wright; William H Schenke; Ozgur Kocaturk; Alexander J Dick; Venkatesh K Raman; Ergin Atalar; Elliot R McVeigh; Robert J Lederman
Journal:  Circulation       Date:  2006-02-20       Impact factor: 29.690

Review 8.  MR-guided endovascular interventions: a comprehensive review on techniques and applications.

Authors:  Sebastian Kos; Rolf Huegli; Georg M Bongartz; Augustinus L Jacob; Deniz Bilecen
Journal:  Eur Radiol       Date:  2007-12-11       Impact factor: 5.315

9.  Closed bore XMR (CBXMR) systems for aortic valve replacement: investigation of rotating-anode x-ray tube heat loadability.

Authors:  John A Bracken; Prasheel V Lillaney; Rebecca Fahrig; J A Rowlands
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

10.  Whole shaft visibility and mechanical performance for active MR catheters using copper-nitinol braided polymer tubes.

Authors:  Ozgur Kocaturk; Christina E Saikus; Michael A Guttman; Anthony Z Faranesh; Kanishka Ratnayaka; Cengizhan Ozturk; Elliot R McVeigh; Robert J Lederman
Journal:  J Cardiovasc Magn Reson       Date:  2009-08-12       Impact factor: 5.364

View more