A new system for the automatic estimation of endothelial cell density in donor corneas.

A Ruggeri1, E Grisan, J Jaroszewski.   

Abstract

AIMS: The problem of automatic estimation of endothelial cell density from microscopy images in donor corneas was addressed.
METHODS: The spatial frequencies contained in digital endothelium images are extracted with a two dimension discrete Fourier transform (DFT) technique. A circular band in the DFT of the images is shown to contain the frequency information related to the cell density. An algorithm for reliably recovering this spatial frequency information and for extracting from it an estimate of endothelial cell density has been developed and implemented in a computer program. An evaluation was performed on a data set containing 100 donor corneas, by comparing automatic values with manual counts performed by three eye bank experts on two images for each cornea.
RESULTS: The mean difference of automatic densities v manual ones was 14 cells/mm(2) (0.9%), with a standard deviation of 119 cells/mm(2) (5.1%) and mean absolute difference of 92 cells/mm(2) (3.9%). The ANOVA based overall inter-rater reliability was 0.935. The algorithm was also capable of identifying all non-processable images. Running times were in the order of 1-2 seconds per image.
CONCLUSION: A new algorithm was developed for the fully automatic estimation of endothelial cell density. The results of a clinical evaluation on 100 corneas suggest that it is capable of reliably estimating endothelium cell density in donor corneas.

Entities:  

Mesh:

Year:  2005        PMID: 15722310      PMCID: PMC1772560          DOI: 10.1136/bjo.2004.051722

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  18 in total

1.  Automated tri-image analysis of stored corneal endothelium.

Authors:  P Gain; G Thuret; L Kodjikian; Y Gavet; P H Turc; C Theillere; S Acquart; J C Le Petit; J Maugery; L Campos
Journal:  Br J Ophthalmol       Date:  2002-07       Impact factor: 4.638

2.  Urgent need for normalization of corneal graft quality controls in French eye banks.

Authors:  Gilles Thuret; Chloé Manissolle; Sophie Acquart; Olivier Garraud; Lydia Campos-Guyotat; Jean Maugery; Philippe Gain
Journal:  Transplantation       Date:  2004-11-15       Impact factor: 4.939

3.  Examination and Photography of donor corneal endothelium.

Authors:  W M Bourne
Journal:  Arch Ophthalmol       Date:  1976-10

4.  Specular microscopic evaluation of donor corneal endothelium.

Authors:  M Matsuda; R W Yee; D B Glasser; D H Geroski; H F Edelhauser
Journal:  Arch Ophthalmol       Date:  1986-02

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Automated analysis of wide-field specular photomicrographs.

Authors:  L W Hirst; R E Sterner; D G Grant
Journal:  Cornea       Date:  1984       Impact factor: 2.651

7.  The corneal endothelium. Normal and pathologic structure and function.

Authors:  G O Waring; W M Bourne; H F Edelhauser; K R Kenyon
Journal:  Ophthalmology       Date:  1982-06       Impact factor: 12.079

8.  Automated morphometric analysis of corneal endothelial cells.

Authors:  J M Lester; J L McFarland; S E Bursell; R A Laing; J F Brenner
Journal:  Invest Ophthalmol Vis Sci       Date:  1981-03       Impact factor: 4.799

9.  Corneal endothelial changes in type I and type II diabetes mellitus.

Authors:  R O Schultz; M Matsuda; R W Yee; H F Edelhauser; K J Schultz
Journal:  Am J Ophthalmol       Date:  1984-10-15       Impact factor: 5.258

View more
  6 in total

1.  Assessment of the human corneal endothelium: in vivo Topcon SP2000P specular microscope versus ex vivo sambacornea eye bank analyser.

Authors:  G Thuret; N Deb-Joardar; C Manissolle; Min Zhao; M Peoch; P Gain; Y Gavet
Journal:  Br J Ophthalmol       Date:  2007-02       Impact factor: 4.638

2.  Evaluation of repeatability for the automatic estimation of endothelial cell density in donor corneas.

Authors:  Alfredo Ruggeri; Enrico Grisan; Jan Schroeter
Journal:  Br J Ophthalmol       Date:  2007-04-12       Impact factor: 4.638

3.  In vivo corneal endothelium imaging using ultrahigh resolution OCT.

Authors:  Xinwen Yao; Kavya Devarajan; René M Werkmeister; Valentin Aranha Dos Santos; Marcus Ang; Anthony Kuo; Damon W K Wong; Jacqueline Chua; Bingyao Tan; Veluchamy Amutha Barathi; Leopold Schmetterer
Journal:  Biomed Opt Express       Date:  2019-10-11       Impact factor: 3.732

4.  Estimating cell count and distribution in labeled histological samples using incremental cell search.

Authors:  Oscar E Meruvia-Pastor; Jung Soh; Eric J Schmidt; Julia C Boughner; Mei Xiao; Heather A Jamniczky; Benedikt Hallgrímsson; Christoph W Sensen
Journal:  Int J Biomed Imaging       Date:  2011-05-24

5.  Fully automatic evaluation of the corneal endothelium from in vivo confocal microscopy.

Authors:  Bettina Selig; Koenraad A Vermeer; Bernd Rieger; Toine Hillenaar; Cris L Luengo Hendriks
Journal:  BMC Med Imaging       Date:  2015-04-26       Impact factor: 1.930

6.  Automatic segmentation of closed-contour features in ophthalmic images using graph theory and dynamic programming.

Authors:  Stephanie J Chiu; Cynthia A Toth; Catherine Bowes Rickman; Joseph A Izatt; Sina Farsiu
Journal:  Biomed Opt Express       Date:  2012-04-26       Impact factor: 3.732

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.