Importance of device evaluation for point-of-care prothrombin time international normalized ratio testing programs.

Robert D McBane1, Cindy L Felty, Mindy L Hartgers, Rajeev Chaudhry, Lisa K Beyer, Paula J Santrach.   

Abstract

OBJECTIVE: To determine the accuracy of 2 commercially available point-of-care devices relative to plasma international normalized ratio (INR) values. PATIENTS AND METHODS: Point-of-care INR testing was performed with the CoaguChek and ProTime 3 devices in consecutive patients attending an anticoagulation clinic between June 18, 2003, and August 6, 2003. Results were compared with plasma INRs using a sensitive thromboplastin (International Sensitivity Index, 1.0).
RESULTS: Ninety-four patients agreed to participate in the study. Relative to the plasma INR, values were in agreement +/-0.4 INR unit 82% and 39% of the time for the CoaguChek and ProTime 3 devices, respectively. The mean +/- SD CoaguChek INRs were 0.2+/-0.31 unit lower, whereas ProTime 3 INRs were 0.8+/-0.68 unit higher than plasma INR values. Treatment decisions based on these data would have resulted in inappropriate dose adjustments 10% and 22% of the time for these 2 respective devices. Correlation with plasma was greater for the CoaguChek (r2=0.90) compared with the ProTime 3 device (r2=0.73).
CONCLUSIONS: Optimal warfarin treatment requires accurate measurement of the INR. The choice of a point-of-care device for INR management depends on the reliability of INR data generated by the device.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15704772     DOI: 10.4065/80.2.181

Source DB:  PubMed          Journal:  Mayo Clin Proc        ISSN: 0025-6196            Impact factor:   7.616


  7 in total

1.  Follow-up and management of valvular heart disease patients with prosthetic valve: a clinical practice guideline for Indian scenario.

Authors:  Devendra Saksena; Yugal K Mishra; S Muralidharan; Vivek Kanhere; Pankaj Srivastava; C P Srivastava
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2019-01-28

2.  In-home Compared With In-Clinic Warfarin Therapy Monitoring in Mechanical Heart Valves: A Population-Based Study.

Authors:  Martin van Zyl; Waldemar E Wysokinski; Thomas M Jaeger; Ana I Casanegra; Bernard J Gersh; Robert D McBane
Journal:  Mayo Clin Proc Innov Qual Outcomes       Date:  2020-08-15

3.  Correlation of point-of-care International Normalized Ratio to laboratory International Normalized Ratio in hemodialysis patients taking warfarin.

Authors:  Robert W Hoel; Robert C Albright; Lisa K Beyer; Paula J Santrach; Donna L Magtibay; Stephanie L Everson; Robert D McBane
Journal:  Clin J Am Soc Nephrol       Date:  2008-11-05       Impact factor: 8.237

4.  Improving anticoagulation therapy using point-of-care testing and a standardized protocol.

Authors:  Curtis A Franke; Lori M Dickerson; Peter J Carek
Journal:  Ann Fam Med       Date:  2008 Jan-Feb       Impact factor: 5.166

5.  Utilization of oral anticoagulation in a teaching hospital in Nigeria.

Authors:  Rc Anakwue; S Ocheni; Aj Madu
Journal:  Ann Med Health Sci Res       Date:  2014-09

6.  Comparison of international normalized ratio measurement between CoaguChek XS Plus and STA-R coagulation analyzers.

Authors:  Mina Hur; Hanah Kim; Chul Min Park; Antonio La Gioia; Sang-Gyu Choi; Ju-Hee Choi; Hee-Won Moon; Yeo-Min Yun
Journal:  Biomed Res Int       Date:  2012-12-24       Impact factor: 3.411

7.  The Effect of Nurse Home-Support Program on Self-Management of Patients Receiving Oral Anticoagulation (Warfarin) Therapy.

Authors:  Jülide Gülizar Yıldırım; Ayla Bayık Temel
Journal:  Florence Nightingale J Nurs       Date:  2020-03-06
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.