Mina Hur1, Hanah Kim, Chul Min Park, Antonio La Gioia, Sang-Gyu Choi, Ju-Hee Choi, Hee-Won Moon, Yeo-Min Yun.
Abstract
BACKGROUND: Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting. We investigated whether the prothrombin time international normalized ratio (INR) results by CoaguChek XS Plus (Roche Diagnostics GmbH, Mannheim, Germany) can be used reliably without being confirmed with the INR results by STA-R system (Diagnostica Stago S.A.S, Asnières sur Seine, France).Entities:
Mesh:
Substances:
Year: 2012 PMID: 23509691 PMCID: PMC3591144 DOI: 10.1155/2013/213109
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1The distribution of prothrombin time (PT) international normalized ratio (INR) measurements by CoaguChek XS Plus and STA-R.
Figure 2Comparison of INR measurements between the CoaguChek XS Plus and the STA-R using Passing/Bablok regression analysis. The solid lines indicate the regression lines, and the dashed lines indicate the 95% confidence interval (CI). (a) In a total of 118 measurements, Passing-Bablok regression analysis gave a slope of 1.22 (95% CI, 1.16–1.28) and an intercept of −0.23 (95% CI = −0.32–0.14). (b) In the range of INR 1.0-2.0 (n = 70), it gave a slope of 1.20 (95% CI, 1.11 to 1.30) and an intercept of −0.21 (95% CI, −0.35 to −0.9). (c) In the range of INR 2.0-3.0 (n = 38), it gave a slope of 1.72 (95% CI, 1.33 to 2.31) and an intercept of −1.36 (95% CI, −2.36 to −0.48).
Figure 3Comparison of INR measurements between the CoaguChek XS Plus and STA-R using Bland-Altman plots in a total of 118 measurements (a), in the range of INR 1.0-2.0 (b), and in the range of INR 2.0-3.0 (c). The difference between two values in the y-axis is plotted against the average of STA-R and CoaguChek XS Plus results in the x-axis. The solid lines represent the mean differences in INR measurements between the two methods, and the dashed lines represent mean difference ±1.96 standard deviation (SD).
Agreement of INR measurements between CoaguChek XS Plus and STA-R.
| CoaguChek XS Plus | Total (%) | |||
|---|---|---|---|---|
| INR < 2 | INR 2.0-3.0 | INR > 3 | ||
| STA-R | ||||
| INR < 2 | 65 | 13 | 0 | 78 (66.1) |
| INR 2.0-3.0 | 1 | 31 | 6 | 38 (32.2) |
| INR > 3.0 | 0 | 1 | 1 | 2 (1.7) |
|
| ||||
| Total (%) | 66 (55.9) | 45 (38.1) | 7 (5.9) | 118 |
Kappa value was 0.679 (95% CI, 0.569–0.790).
Previous studies on the performances of CoaguChek XS Plus system.
| Study | Number of measurements | Setting | Laboratory-based system | Comparison using Bland-Altman plot, |
|---|---|---|---|---|
| Deom et al. (2009) [ | 259 | Community setting | Innovin (Siemens/Dade-Behring) | 0.03 INR (−0.67 INR ~ 0.74 INR) |
| Urwyler et al. (2009) [ | 227 | Perioperative setting | Innovin (Siemens/Dade-Behring) | 2.3% (95% CI = −20.3% ~ 24.9%) |
| Donaldson et al. (2010) [ | 52 | Pharmacist-managed anticoagulation clinic | Stago (Diagnostica-stago) | 0.27 INR (−0.33 INR ~ 0.88 INR) |
|
Celenza and Skinner (2011) [ | 293 | Emergency department | Not specified | −0.0 INR (−1.1 INR ~ 1.1 INR) |
| Lawrie et al. (2012) [ | 168 for innovin and 115 for PT-Fib HS+ | Overanticoagulated (INR > 4.5) patients in anticoagulation clinic | Innovin (Siemens/Dade-Behring) on CA-7000 analyzer (Sysmex) and PT-Fib HS+ (Instrumentation Laboratory) on CA-1500 analyzer (Sysmex) | Innovin: |
| Urwyler et al. (2012) [ | 73 | Pediatric intensive care unit setting | Innovin (Siemens/Dade-Behring) | 1.22% (−27% ~ 56%) |
Abbreviations: INR: international normalized ratio; SD: standard deviation; CI: confidence interval; PT-Fib HS+: HemosIL PT-Fibrinogen HS Plus.