Evaluation of Roche Accu-Chek Go and Medisense Optium blood glucose meters.

Robert C Hawkins1.   

Abstract

BACKGROUND: We examined the analytical performance of two Roche Accu-Chek Go meters (reporting whole blood glucose) and two Medisense Optium meters (reporting plasma glucose) against laboratory plasma glucose measurement.
METHODS: One hundred and twenty heparinized whole blood specimens were analyzed in duplicate on each of the four meters and compared to plasma glucose measurement on the Roche DP Modular system (glucose oxidase methodology).
RESULTS: There was a significant difference in imprecision between the two Medisense meters (CV 3.7% and 4.9%), and between the Roche (CV 2.5% and 2.9%) and Medisense meters. The two Roche meters showed a consistent positive bias compared to the Modular, while the two Medisense meters over-read at low glucose concentrations and under-read at high glucose concentrations. Total error for the Roche meters averaged 10.8% and 8.9%, and for the Medisense meters averaged 12.5% and 15.6%, respectively. All results for all meters fell in error grid zones A and B.
CONCLUSION: Neither meter met the ADA, CLIA or biological variation goals, although they came close to the NCCLS guideline. Neither meter reported plasma glucose equivalents. The Accu-Chek Go performed slightly better than the Optium meter, but the positive bias was puzzling and requires further explanation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15698599     DOI: 10.1016/j.cccn.2004.10.013

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  6 in total

Review 1.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Evaluation of the analytical performance of the coulometry-based Optium Omega blood glucose meter.

Authors:  Bogdan Solnica; Beata Kusnierz-Cabala; Krystyna Slowinska-Solnica; Przemyslaw Witek; Agnieszka Cempa; Maciej T Malecki
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Quality control of self-monitoring of blood glucose: why and how?

Authors:  Bogdan Solnica; Jerzy W Naskalski
Journal:  J Diabetes Sci Technol       Date:  2007-03

4.  m-Health: Lessons Learned by m-Experiences.

Authors:  José Bravo; Ramón Hervás; Jesús Fontecha; Iván González
Journal:  Sensors (Basel)       Date:  2018-05-15       Impact factor: 3.576

5.  Cafeteria Diet and High-Fructose Rodent Models of NAFLD Differ in the Metabolism of Important PUFA and Palmitoleic Acid without Additional Influence of Sex.

Authors:  Tomislav Mašek; Josip Barišić; Vedran Micek; Kristina Starčević
Journal:  Nutrients       Date:  2020-10-30       Impact factor: 5.717

6.  Low Dietary n6/n3 Ratio Attenuates Changes in the NRF 2 Gene Expression, Lipid Peroxidation, and Inflammatory Markers Induced by Fructose Overconsumption in the Rat Abdominal Adipose Tissue.

Authors:  Petra Roškarić; Marcela Šperanda; Tomislav Mašek; Donatella Verbanac; Kristina Starčević
Journal:  Antioxidants (Basel)       Date:  2021-12-16
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.