Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error?

Yat-Yin Yau1, Wing-Sze Chan, Yat-Man Tam, Phil Vernon, So Wong, Marc Coel, Simon Kwok-Fai Chu.   

Abstract

UNLABELLED: The current perception of using contrast-enhanced CT (CECT) for attenuation correction (AC) is that of caution, as it might lead to erroneously elevated (18)F-FDG uptake on the PET scan. This study evaluates in vivo whether an intravenous iodinated contrast agent produces a significant AC artifact in the level of standardized uptake value (SUV) changes in PET/CT.
METHODS: Fifty-four patients referred for whole-body (WB) PET/CT scans were enrolled and subdivided into 2 groups. In part I, 26 patients had a single WB PET scan that was corrected for attenuation using noncontrast and intravenous CECT obtained before and after the emission data, respectively. The final PET images were compared for any visual and SUV maximum (SUV(max)) measurement difference. This allowed analysis of the compatibility of the scaling processes between the 2 different CTs and the PET. The SUV(max) values were obtained from ascending aorta, upper lung, femoral head, iliopsoas muscle, spleen, liver, and the site of pathology (total, 193 regions). Part II addressed whether intravenous contrast also influenced the PET emission data. For that purpose, the remaining 28 patients underwent a limited plain CT scan from lung base to lower liver edge, followed by a 1-bed PET scan of the same region and then a WB intravenous contrast CT scan in tandem with a WB PET scan. SUV(max) values were obtained at the lung base, liver, spleen, T11 or T12 vertebra, and paraspinal muscle (total, 135 regions). The data obtained from pre- and post-intravenous contrast PET scans were analyzed as in part I.
RESULTS: There was no statistically significant elevation of the SUV level in the measured anatomic sites as a whole (part I: mean SUV(max) difference = 0.06, P > 0.05; Part II: mean SUV(max) difference = -0.02, P > 0.05). However, statistically significant results as a group (mean SUV(max) difference = 0.26, P < 0.05)--albeit considered to be clinically insignificant--were observed for areas of pathology in the part I study. No abnormal focal increased (18)F-FDG activity was detected as a result of the intravenous contrast in both parts of this examination.
CONCLUSION: No statistically or clinically significant spuriously elevated SUV level that might potentially interfere with the diagnostic value of PET/CT was identified as a result of the application of intravenous iodinated contrast.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15695788

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  30 in total

1.  PET/CT in lung cancer: Influence of contrast medium on quantitative and clinical assessment.

Authors:  Florian F Behrendt; Yavuz Temur; Frederik A Verburg; Moritz Palmowski; Thomas Krohn; Hubertus Pietsch; Christiane K Kuhl; Felix M Mottaghy
Journal:  Eur Radiol       Date:  2012-06-04       Impact factor: 5.315

2.  Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging.

Authors:  Philip Aschoff; Christian Plathow; Thomas Beyer; Matthias P Lichy; Gunter Erb; Mehmet Ö Öksüz; Claus D Claussen; Christina Pfannenberg
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-11-29       Impact factor: 9.236

3.  Calcified Lymph Nodes Causing Clinically Relevant Attenuation Correction Artifacts on PET/CT Imaging.

Authors:  Amar Mehta; Ajeet Mehta; Charles Laymon; Todd M Blodgett
Journal:  J Radiol Case Rep       Date:  2010-02-01

4.  Assessment of errors caused by X-ray scatter and use of contrast medium when using CT-based attenuation correction in PET.

Authors:  Mohammad Reza Ay; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-04-19       Impact factor: 9.236

5.  Additional value of PET-CT in the staging of lung cancer: comparison with CT alone, PET alone and visual correlation of PET and CT.

Authors:  W De Wever; S Ceyssens; L Mortelmans; S Stroobants; G Marchal; J Bogaert; J A Verschakelen
Journal:  Eur Radiol       Date:  2006-05-09       Impact factor: 5.315

6.  Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements.

Authors:  Tira Bunyaviroch; Timothy G Turkington; Terence Z Wong; John W Wilson; James G Colsher; R Edward Coleman
Journal:  Mol Imaging Biol       Date:  2007-12-21       Impact factor: 3.488

7.  The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials.

Authors:  Ronald Boellaard; Wim J G Oyen; Corneline J Hoekstra; Otto S Hoekstra; Eric P Visser; Antoon T Willemsen; Bertjan Arends; Fred J Verzijlbergen; Josee Zijlstra; Anne M Paans; Emile F I Comans; Jan Pruim
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-08-15       Impact factor: 9.236

8.  Comparison of integrated whole-body [11C]choline PET/MR with PET/CT in patients with prostate cancer.

Authors:  Michael Souvatzoglou; Matthias Eiber; Toshiki Takei; Sebastian Fürst; Tobias Maurer; Florian Gaertner; Hans Geinitz; Alexander Drzezga; Sibylle Ziegler; Stephan G Nekolla; Ernst J Rummeny; Markus Schwaiger; Ambros J Beer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-07-02       Impact factor: 9.236

9.  Correction of oral contrast artifacts in CT-based attenuation correction of PET images using an automated segmentation algorithm.

Authors:  Alireza Ahmadian; Mohammad R Ay; Javad H Bidgoli; Saeed Sarkar; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-04-17       Impact factor: 9.236

Review 10.  PET/CT in oncology: for which tumours is it the reference standard?

Authors:  Conor D Collins
Journal:  Cancer Imaging       Date:  2007-10-01       Impact factor: 3.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.