Accurate mass measurement in nano-electrospray ionization mass spectrometry by alternate switching of high voltage between sample and reference sprayers.

Yoshinori Satomi1, Yoichiro Kudo, Kyoko Sasaki, Toshiharu Hase, Toshifumi Takao.   

Abstract

An electrospray dual sprayer, which generates separate sample and reference sprays by alternately switching the high voltage between the two sprayers, is described. The technique permits accurate mass measurements in nano-electrospray ionization mass spectrometry (ESI-MS) to be obtained using a quadrupole/orthogonal acceleration time-of-flight mass spectrometer (Q-TOF). Similar to the method employed with a dual ESI source (Wolff JC et al., Anal. Chem. 2001; 73: 2605), the two sprays are orthogonal with respect to each other, but can be independently sampled without any baffle between these sprays. The reference sprayer is used in the original configuration of the ESI source and was optimized for a 1-2 muL/min flow, whereas the sample sprayer can be either a conventional glass capillary or a borosilicate tip of the type used for nano-ESI. Both sprayers can be positioned close to the cone so as to give maximum ion currents. The sample and reference sprays are independently generated by raising the potentials on the sample and reference sprayers to 1.4 and 3.0 kV, respectively; the high voltages can be rapidly turned on and off in ca. 1 ms. A nano-ESI-MS or nano-flow LC/ESI-MS experiment using a Q-TOF coupled with the above system gave mass accuracies within 3 ppm for measurements of ions up to m/z 1000 using subpicomole samples. Copyright 2005 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15666318     DOI: 10.1002/rcm.1813

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  5 in total

1.  Steady-state asymmetric nanospray dual ion source for accurate mass determination within a chromatographic separation.

Authors:  Nicolas L Young; Michael C Sisto; Meggie N Young; Patrick G Grant; David W Killilea; LaTasha LaMotte; Kuang Jen J Wu; Carlito B Lebrilla
Journal:  Anal Chem       Date:  2007-07-06       Impact factor: 6.986

2.  Induced dual-nanospray: a novel internal calibration method for convenient and accurate mass measurement.

Authors:  Yafeng Li; Ning Zhang; Yueming Zhou; Jianing Wang; Yiming Zhang; Jiyun Wang; Caiqiao Xiong; Suming Chen; Zongxiu Nie
Journal:  J Am Soc Mass Spectrom       Date:  2013-09       Impact factor: 3.109

3.  Pulsed dual electrospray ionization for ion/ion reactions.

Authors:  Yu Xia; Xiaorong Liang; Scott A McLuckey
Journal:  J Am Soc Mass Spectrom       Date:  2005-09-22       Impact factor: 3.109

4.  Pulsed Nano-Electrospray Ionization: Characterization of Temporal Response and Implementation with a Flared Inlet Capillary.

Authors:  Jared M Bushey; Desmond A Kaplan; Ryan M Danell; Gary L Glish
Journal:  Instrum Sci Technol       Date:  2009       Impact factor: 1.584

5.  Multi-channel microfluidic chip coupling with mass spectrometry for simultaneous electro-sprays and extraction.

Authors:  Cilong Yu; Fei Tang; Xiang Qian; Yan Chen; Quan Yu; Kai Ni; Xiaohao Wang
Journal:  Sci Rep       Date:  2017-12-12       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.