OBJECTIVE: Prostate-specific antigen (PSA) samples from five commercial sources were compared with PSA from the European Commission Community Bureau of Reference, known as CRM 613, using liquid chromatography-mass spectrometry (LC-MS) and matrix-assisted laser desorption ionization-mass spectrometry (MALDI-MS). DESIGN AND METHODS: The effects of storage at various temperatures, multiple freeze-thaw cycles and lyophilization vs. solution storage were investigated.
RESULTS: : CRM 613 and the five commercially available samples varied in their heterogeneity and in the amount of PSA with a molecular mass of 28,430 kDa. Few changes were observed among any of the samples during the stability study or study of the effect of lyophilization and freeze-thaw cycles.
CONCLUSIONS: Use of many of the commercial PSA sources as authentic PSA standards for calibrants for in vitro diagnostic tests could lead to significantly biased PSA measurements. Although all the PSA samples appear to be relatively stable for the time periods and conditions studied, some of the commercially available samples contain limited amounts of PSA and included other unidentified components.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15642280     DOI: 10.1016/j.clinbiochem.2004.10.016

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  5 in total

Review 1.  Mass spectrometry and the emerging field of glycomics.

Authors:  Joseph Zaia
Journal:  Chem Biol       Date:  2008-09-22

2.  Reference materials and commutability.

Authors:  Hubert W Vesper; W Gregory Miller; Gary L Myers
Journal:  Clin Biochem Rev       Date:  2007-11

Review 3.  Mass spectrometry and glycomics.

Authors:  Joseph Zaia
Journal:  OMICS       Date:  2010-08

4.  Influence of PEG coating on the biodistribution and tumor accumulation of pH-sensitive liposomes.

Authors:  Shirleide Santos Nunes; Renata Salgado Fernandes; Carolina Henriques Cavalcante; Isabela da Costa César; Elaine Amaral Leite; Sávia Caldeira Araújo Lopes; Alice Ferretti; Domenico Rubello; Danyelle M Townsend; Mônica Cristina de Oliveira; Valbert Nascimento Cardoso; André Luís Branco de Barros
Journal:  Drug Deliv Transl Res       Date:  2019-02       Impact factor: 4.617

Review 5.  Review of the Use of Liquid Chromatography-Tandem Mass Spectrometry in Clinical Laboratories: Part II-Operations.

Authors:  Brian A Rappold
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.