Abstract

BACKGROUND: The purpose of this study was to assess the value of fluoroscopy-based computer-assisted surgery (CAS) for the insertion of iliosacral screws. The results of CAS were compared with the results of a conventionally operated prospective control group. Endpoints of this study were fluoroscopy time, guide wire insertion time, operation time and complication rate.
METHODS: The study group consisted of 24 patients with postpartum pelvic pain syndrome. All patients were treated with a stabilization of the pelvic ring by means of an anterior plate fixation and autologous tricortical bone graft as well as two iliosacral screws bilaterally. Consequently, the results of 48 versus 48 iliosacral screw fixations could be evaluated. Conventionally operated patients were turned from the supine to the prone position intraoperatively, whereas CAS operated patients were operated in the supine position. One surgeon performed all operations.
RESULTS: The fluoroscopy time in the CAS group was 0.7 minutes versus 1.8 minutes in the conventionally treated group (p < 0.01). The mean insertion time for four guide wires was 20.2 minutes in the CAS versus 19.4 minutes in the conventionally operated group (p = 0.6). The mean operation time in the CAS group was 97 minutes; 116 minutes in the conventional group (p = 0.03). In the CAS group one patient had pain and a sensory deficit of S2 postoperatively. The Fisher's exact test showed no difference in complication rate between the two groups (p = 0.26).
CONCLUSIONS: The fluoroscopy time is decreased with a factor 2.5 using CAS. Guide wire insertion time was similar in both groups. The reduction in operation time using CAS was due to fact that patients were operated in the supine position during the whole procedure. This study shows that CAS is a save technique for insertion of iliosacral screws.

Entities:  

Mesh:

Year:  2004        PMID: 15625463     DOI: 10.1097/01.ta.0000133573.53587.2e

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  22 in total

1.  Factors influencing the accuracy of iliosacral screw placement in trauma patients.

Authors:  Lars Grossterlinden; Johannes Rueger; Philip Catala-Lehnen; Martin Rupprecht; Wolfgang Lehmann; Andreas Rücker; Daniel Briem
Journal:  Int Orthop       Date:  2010-07-20       Impact factor: 3.075

2.  [Computer-assisted arthrodesis of the sacroiliac joint].

Authors:  F Gebhard; M Sarkar; G Maier; L Kinzl; M Arand
Journal:  Unfallchirurg       Date:  2005-09       Impact factor: 1.000

3.  Computer-assisted screw insertion into the first sacral vertebra using a three-dimensional image intensifier: results of a controlled experimental investigation.

Authors:  D Briem; W Linhart; W Lehmann; P G Begemann; G Adam; U Schumacher; D M Cullinane; J M Rueger; J Windolf
Journal:  Eur Spine J       Date:  2005-07-12       Impact factor: 3.134

4.  [Computer-assisted screw placement into the posterior pelvic ring: assessment of different navigated procedures in a cadaver trial].

Authors:  D Briem; J M Rueger; P G C Begemann; Z Halata; T Bock; W Linhart; J Windolf
Journal:  Unfallchirurg       Date:  2006-08       Impact factor: 1.000

5.  [Percutaneous, 2D-fluoroscopic navigated iliosacral screw placement in the supine position: technique, possibilities, and limits].

Authors:  D Briem; J Windolf; J M Rueger
Journal:  Unfallchirurg       Date:  2007-05       Impact factor: 1.000

6.  [Bony sacroiliac corridor. A virtual volume model for the accurate insertion of transarticular screws].

Authors:  T Mendel; K Appelt; P Kuhn; N Suhm
Journal:  Unfallchirurg       Date:  2008-01       Impact factor: 1.000

7.  Computer-navigated iliosacral screw insertion reduces malposition rate and radiation exposure.

Authors:  Jörn Zwingmann; Gerhard Konrad; Elmar Kotter; Norbert P Südkamp; Michael Oberst
Journal:  Clin Orthop Relat Res       Date:  2008-11-26       Impact factor: 4.176

Review 8.  Radiation protection and standardization.

Authors:  O P Lakhwani; Vipin Dalal; Mohit Jindal; Ashok Nagala
Journal:  J Clin Orthop Trauma       Date:  2018-08-07

9.  Can computer-assisted surgery reduce the effective dose for spinal fusion and sacroiliac screw insertion?

Authors:  Michael David Kraus; Gert Krischak; Peter Keppler; Florian T Gebhard; Uwe H W Schuetz
Journal:  Clin Orthop Relat Res       Date:  2010-06-03       Impact factor: 4.176

10.  2D-fluoroscopic navigated percutaneous screw fixation of pelvic ring injuries--a case series.

Authors:  Florian Gras; Ivan Marintschev; Arne Wilharm; Kajetan Klos; Thomas Mückley; Gunther O Hofmann
Journal:  BMC Musculoskelet Disord       Date:  2010-07-07       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.