Literature DB >> 15617932

Analysis of the capsule edge after Fugo plasma blade capsulotomy, continuous curvilinear capsulorhexis, and can-opener capsulotomy.

Andrea M Izak1, Liliana Werner, Suresh K Pandey, David J Apple, Milan G J Izak.   

Abstract

PURPOSE: To compare the quality of the edges of anterior capsulotomies performed in porcine eyes using 3 different techniques.
SETTING: David J. Apple, MD Laboratories for Ophthalmic Devices Research, John A. Moran Eye Center, Department of Ophthalmology and Visual Sciences, University of Utah, Salt Lake City, Utah, USA.
METHODS: Four porcine eyes were submitted to the center in 10% neutral buffered formalin. The cornea of the eyes had been removed, and a continuous capsulotomy had been performed in each eye using the Fugo plasma blade. To provide morphologic correlation, 4 additional porcine eyes were used in the study. In 2, a manual continuous curvilinear capsulorhexis (CCC) was performed after the cornea or cornea/iris was removed and in the other 2, a can-opener capsulotomy (COC) was performed. The eyes were then placed in 10% neutral buffered formalin. Gross (macroscopic) and microscopic analysis of each eye was performed, and photographs were taken. One eye in each group was also prepared and analyzed under scanning electron microscopy (SEM).
RESULTS: The capsulotomy margin in the Fugo blade group presented some regularly spaced tags directed centrally with smooth edges at the base. Such a configuration would not interfere with the dynamics of the capsulorhexis sphincter and would generally not lead to radial tears. The smooth and regular nature of the capsulotomy margin in the CCC group was confirmed by gross and microscopic examinations of the eyes. The capsulotomy margin in the COC group presented some sharp notches directed outward, irregularly spaced, and irregular in shape. These are known to have a tendency to extend and form radial tears.
CONCLUSION: Although performance of an anterior capsulotomy with the Fugo blade was associated with some margin irregularities, the geometry of the centrally directed tags prevented them from becoming the site of radial tear formation. The base of the capsulotomy performed with the Fugo blade appeared to be almost as smooth and regular as in the CCC under SEM evaluation.

Entities:  

Mesh:

Year:  2004        PMID: 15617932     DOI: 10.1016/j.jcrs.2004.05.020

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Pulsed electron avalanche knife: new technology for cataract surgery.

Authors:  Siegfried G Priglinger; Daniel Palanker; Claudia S Alge; Thomas C Kreutzer; Christos Haritoglou; Martin Grueterich; Anselm Kampik
Journal:  Br J Ophthalmol       Date:  2007-01-17       Impact factor: 4.638

2.  Capsulorhexis: Pearls and pitfalls.

Authors:  Mehrdad Mohammadpour; Reza Erfanian; Nasser Karimi
Journal:  Saudi J Ophthalmol       Date:  2012-01

3.  Anterior capsulotomy with a pulsed-electron avalanche knife.

Authors:  Daniel Palanker; Hiroyuki Nomoto; Philip Huie; Alexander Vankov; David F Chang
Journal:  J Cataract Refract Surg       Date:  2010-01       Impact factor: 3.351

4.  Results of cataract surgery and plasma ablation posterior capsulotomy in anterior persistent hyperplastic primary vitreous.

Authors:  Rajesh Sinha; Shveta Jindal Bali; Chandrashekhar Kumar; Himanshu Shekhar; Namrata Sharma; Jeewan S Titiyal; Rasik B Vajpayee
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

5.  Scanning Electron Microscopy Analysis of the Anterior Capsulotomy Edge: A Comparative Study between Femtosecond Laser-Assisted Capsulotomy and Manual Capsulorhexis.

Authors:  Daniele Tognetto; Chiara De Giacinto; Alberto Armando Perrotta; Tommaso Candian; Alessandro Bova; Silvia Rinaldi; Gianluca Turco
Journal:  J Ophthalmol       Date:  2018-11-14       Impact factor: 1.909

Review 6.  Techniques of anterior capsulotomy in cataract surgery.

Authors:  Bhavana Sharma; Robin G Abell; Tarun Arora; Tom Antony; Rasik B Vajpayee
Journal:  Indian J Ophthalmol       Date:  2019-04       Impact factor: 1.848

7.  Modified bandage-contact-lens used as a guide-marker for performing continuous-curvilinear-capsulorhexis by a first-year-post-graduate-ophthalmology-resident.

Authors:  Mayuresh P Naik; HarinderSingh Sethi; Pulkit Kasiviswanathan
Journal:  Am J Ophthalmol Case Rep       Date:  2020-08-26

8.  Methods of advanced wound management for care of combined traumatic and chemical warfare injuries.

Authors:  John S Graham; Travis W Gerlach; Thomas P Logan; James P Bonar; Richard J Fugo; Robyn B Lee; Matthew A Coatsworth
Journal:  Eplasty       Date:  2008-07-21
  8 in total