A liquid chromatographic-mass spectrometric assay with atmospheric pressure chemical ionization (LC-APCI-MS) is presented for screening for, library-assisted identification (both in scan mode) and quantification (selected-ion mode) of the beta-blockers acebutolol, diacetolol, alprenolol, atenolol, betaxolol, bisoprolol, bupranolol, carazolol, carteolol, carvedilol, celiprolol, esmolol, labetalol, metoprolol, nadolol, nebivolol, oxprenolol, penbutolol, propranolol, sotalol, talinolol and timolol in blood plasma after mixed-mode (HCX) solid-phase extraction (SPE) and separation by reverse-phase liquid chromatography with gradient elution. The validation data were within the required limits. The assay was successfully applied to authentic plasma samples allowing confirmation of diagnosis of overdose situations as well as monitoring of patients' compliance.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15595665