Performance of a commercial optical CT scanner and polymer gel dosimeters for 3-D dose verification.

Y Xu1, Cheng-Shie Wuu, Marek J Maryanski.   

Abstract

Performance analysis of a commercial three-dimensional (3-D) dose mapping system based on optical CT scanning of polymer gels is presented. The system consists of BANG 3 polymer gels (MGS Research, Inc., Madison, CT), OCTOPUS laser CT scanner (MGS Research, Inc., Madison, CT), and an in-house developed software for optical CT image reconstruction and 3-D dose distribution comparison between the gel, film measurements and the radiation therapy treatment plans. Various sources of image noise (digitization, electronic, optical, and mechanical) generated by the scanner as well as optical uniformity of the polymer gel are analyzed. The performance of the scanner is further evaluated in terms of the reproducibility of the data acquisition process, the uncertainties at different levels of reconstructed optical density per unit length and the effects of scanning parameters. It is demonstrated that for BANG 3 gel phantoms held in cylindrical plastic containers, the relative dose distribution can be reproduced by the scanner with an overall uncertainty of about 3% within approximately 75% of the radius of the container. In regions located closer to the container wall, however, the scanner generates erroneous optical density values that arise from the reflection and refraction of the laser rays at the interface between the gel and the container. The analysis of the accuracy of the polymer gel dosimeter is exemplified by the comparison of the gel/OCT-derived dose distributions with those from film measurements and a commercial treatment planning system (Cadplan, Varian Corporation, Palo Alto, CA) for a 6 cm x 6 cm single field of 6 MV x rays and a 3-D conformal radiotherapy (3DCRT) plan. The gel measurements agree with the treatment plans and the film measurements within the "3%-or-2 mm" criterion throughout the usable, artifact-free central region of the gel volume. Discrepancies among the three data sets are analyzed.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15587656     DOI: 10.1118/1.1803674

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  A practical three-dimensional dosimetry system for radiation therapy.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

2.  Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation.

Authors:  H S Sakhalkar; M Oldham
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  Quality assurance in 3D dosimetry by optical-CT.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006-12-01

4.  A dual-purpose CCD based micro-optical-CT scanning system.

Authors:  M Oldham; H Sakhalkar; P Guo
Journal:  J Phys Conf Ser       Date:  2006-12-01

5.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

6.  A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system.

Authors:  H S Sakhalkar; J Adamovics; G Ibbott; M Oldham
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

7.  A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique.

Authors:  Andrew Thomas; Michael Niebanck; Titania Juang; Zhiheng Wang; Mark Oldham
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

8.  Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

Authors:  Harshad Sakhalkar; David Sterling; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

9.  Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters.

Authors:  Y Xu; Cheng-Shie Wuu; Marek J Maryanski
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

10.  Commissioning and benchmarking a 3D dosimetry system for clinical use.

Authors:  Andrew Thomas; Joseph Newton; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.