Randomized clinical trial of extended use of a hydrophobic condenser humidifier: 1 vs. 7 days.

Laurent Thomachot1, Marc Leone, Karim Razzouk, François Antonini, Renaud Vialet, Claude Martin.   

Abstract

OBJECTIVE: To determine whether extended use (7 days) would affect the efficiency on heat and water preservation of a hydrophobic condenser humidifier as well as the rate of ventilation-acquired pneumonia, compared with 1 day of use.
DESIGN: Prospective, controlled, randomized, not blinded, clinical study.
SETTING: Twelve-bed intensive care unit of a university hospital. PATIENTS: One hundred and fifty-five consecutive patients undergoing mechanical ventilation for > or = 48 hrs.
INTERVENTIONS: After randomization, patients were allocated to one of the two following groups: a) heat and moisture exchangers (HMEs) changed every 24 hrs; b) HMEs changed only once a week. Devices in both groups could be changed at the discretion of the staff when signs of occlusion or increased resistance were identified. MEASUREMENTS AND MAIN RESULT: Efficient airway humidification and heating were assessed by clinical variables (numbers of tracheal suctionings and instillations required, peak and mean airway pressures). The frequency rates of bronchial colonization and ventilation-acquired pneumonia were evaluated by using clinical and microbiological criteria. Endotracheal tube occlusion, ventilatory support variables, duration of mechanical ventilation, length of intensive care, acquired multiorgan dysfunction, and mortality rates also were recorded. The two groups were similar at the time of randomization. Endotracheal tube occlusion never occurred. In the targeted population (patients ventilated for > or = 7 days), the frequency rate of ventilation-acquired pneumonia was 24% in the HME 1-day group and 17% in the HME 7-day group (p > .05, not significant). Ventilation-acquired pneumonia rates per 1000 ventilatory support days were 16.4/1000 in the HME 1-day group and 12.4/1000 in the HME 7-day group (p > .05, not significant). No statistically significant differences were found between the two groups for duration of mechanical ventilation, intensive care unit length of stay, acquired organ system derangements, and mortality rate. There was indirect evidence of very little, if any, change in HME resistance.
CONCLUSIONS: Changing the studied hydrophobic HME after 7 days did not affect efficiency, increase resistance, or altered bacterial colonization. The frequency rate of ventilation-acquired pneumonia was also unchanged. Use of HMEs for > 24 hrs and up to 7 days is safe.

Entities:  

Mesh:

Year:  2002        PMID: 11902268     DOI: 10.1097/00003246-200201000-00033

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  13 in total

Review 1.  Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials.

Authors:  Axel Kola; Tim Eckmanns; Petra Gastmeier
Journal:  Intensive Care Med       Date:  2004-09-11       Impact factor: 17.440

2.  Airway humidification with a heat and moisture exchanger in mechanically ventilated neonates : a preliminary evaluation.

Authors:  Mikaïla Fassassi; Fabrice Michel; Laurent Thomachot; Claire Nicaise; Renaud Vialet; Yves Jammes; Pierre Lagier; Claude Martin
Journal:  Intensive Care Med       Date:  2006-12-13       Impact factor: 17.440

3.  Ventilator-associated pneumonia: diagnosis, treatment, and prevention.

Authors:  Steven M Koenig; Jonathon D Truwit
Journal:  Clin Microbiol Rev       Date:  2006-10       Impact factor: 26.132

4.  Comparison of the effects of two humidifier systems on endotracheal tube resistance.

Authors:  Indalecio Morán; Belén Cabello; Elizabeth Manero; Jordi Mancebo
Journal:  Intensive Care Med       Date:  2011-09-27       Impact factor: 17.440

5.  Endotracheal tube intraluminal diameter narrowing after mechanical ventilation: use of acoustic reflectometry.

Authors:  M C Boqué; B Gualis; A Sandiumenge; J Rello
Journal:  Intensive Care Med       Date:  2004-10-02       Impact factor: 17.440

Review 6.  Humidification and heating of inhaled gas in patients with artificial airway. A narrative review.

Authors:  Gustavo Adrián Plotnikow; Matias Accoce; Emiliano Navarro; Norberto Tiribelli
Journal:  Rev Bras Ter Intensiva       Date:  2018-03

7.  Effectiveness of heat and moisture exchangers in preventing ventilator-associated pneumonia in critically ill patients: a meta-analysis.

Authors:  Mayra Gonçalves Menegueti; Maria Auxiliadora-Martins; Altacílio Aparecido Nunes
Journal:  BMC Anesthesiol       Date:  2014-12-13       Impact factor: 2.217

Review 8.  Humidification during mechanical ventilation in the adult patient.

Authors:  Haitham S Al Ashry; Ariel M Modrykamien
Journal:  Biomed Res Int       Date:  2014-06-25       Impact factor: 3.411

9.  Effect of heat and moisture exchangers on the prevention of ventilator-associated pneumonia in critically ill patients.

Authors:  M Auxiliadora-Martins; M G Menegueti; E A Nicolini; G C Alkmim-Teixeira; F Bellissimo-Rodrigues; O A Martins-Filho; A Basile-Filho
Journal:  Braz J Med Biol Res       Date:  2012-10-09       Impact factor: 2.590

10.  Airway accidents in critical care unit: A 3-year retrospective study in a Public Teaching Hospital of Eastern India.

Authors:  Sugata Dasgupta; Shipti Shradha Singh; Arunima Chaudhuri; Dipasri Bhattacharya; Sourav Das Choudhury
Journal:  Indian J Crit Care Med       Date:  2016-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.