Literature DB >> 15578566

A hip joint simulator study of the performance of metal-on-metal joints: Part II: design.

Duncan Dowson1, Cath Hardaker, Magnus Flett, Graham H Isaac.   

Abstract

The separate and combined roles of head diameter and clearance were studied experimentally in simulator tests and also theoretically. Head diameters ranging from 16-54 mm and various clearances were studied. Effective mixed-film lubrication achieved through careful design and manufacture greatly reduced wear. The running-in wear volumes were very low for larger-diameter heads and smaller clearances; but because the subsequent steady-state wear rate was much smaller than the running-in wear rate, the former contributed significantly to the total volume of wear debris even over very long periods of time. The design study has shown that head diameters should be as large as possible and diametral clearances as low as practicable, to ensure that the joints operate well into the mixed lubrication regime.

Entities:  

Mesh:

Year:  2004        PMID: 15578566     DOI: 10.1016/j.arth.2004.09.016

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  26 in total

1.  Wear degradation of long-term in vivo exposed alumina-on-alumina hip joints: linking nanometer-scale phenomena to macroscopic joint design.

Authors:  Yasuhito Takahashi; Nobuhiko Sugano; Wenliang Zhu; Takashi Nishii; Takashi Sakai; Masaki Takao; Giuseppe Pezzotti
Journal:  J Mater Sci Mater Med       Date:  2011-11-22       Impact factor: 3.896

2.  Ten different hip resurfacing systems: biomechanical analysis of design and material properties.

Authors:  Christian Heisel; Jennifer A Kleinhans; Michael Menge; Jan Philippe Kretzer
Journal:  Int Orthop       Date:  2008-07-04       Impact factor: 3.075

3.  Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  Blood metal ion concentrations in metal-on-metal total hip arthroplasty.

Authors:  Tadahiko Ohtsuru; Yuji Morita; Yasuaki Murata; Shuji Shimamoto; Yutaro Munakata; Yoshiharu Kato
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-02-20

5.  What are the risks accompanying the reduced wear benefit of low-clearance hip resurfacing?

Authors:  Joseph Daniel; Hena Ziaee; Amir Kamali; Chandra Pradhan; Derek McMinn
Journal:  Clin Orthop Relat Res       Date:  2012-07-24       Impact factor: 4.176

6.  2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial.

Authors:  C Anderson Engh; Steven J MacDonald; Supatra Sritulanondha; Abigail Thompson; Douglas Naudie; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2008-10-15       Impact factor: 4.176

7.  The natural history of inflammatory pseudotumors in asymptomatic patients after metal-on-metal hip arthroplasty.

Authors:  Sulaiman A Almousa; Nelson V Greidanus; Bassam A Masri; Clive P Duncan; Donald S Garbuz
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

8.  What Is the Natural History of Asymptomatic Pseudotumors in Metal-on-metal THAs at Mid-term Followup?

Authors:  Sujith Konan; Clive P Duncan; Bassam S Masri; Donald S Garbuz
Journal:  Clin Orthop Relat Res       Date:  2017-02       Impact factor: 4.176

9.  What are the predictors and prevalence of pseudotumor and elevated metal ions after large-diameter metal-on-metal THA?

Authors:  Nick Bayley; Habeeb Khan; Paul Grosso; Thomas Hupel; David Stevens; Matthew Snider; Emil Schemitsch; Paul Kuzyk
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

10.  Influence of head size on the development of metallic wear and on the characteristics of carbon layers in metal-on-metal hip joints.

Authors:  Volker Braunstein; Christoph M Sprecher; Markus A Wimmer; Stefan Milz; Georg Taeger
Journal:  Acta Orthop       Date:  2009-06       Impact factor: 3.717

View more