Measurement and correction of respiration-induced B0 variations in breast 1H MRS at 4 Tesla.

Patrick J Bolan1, Pierre-Gilles Henry, Eva H Baker, Sina Meisamy, Michael Garwood.   

Abstract

Respiratory motion is well known to cause artifacts in magnetic resonance spectroscopy (MRS). In MRS of the breast, the dominant artifact is not due to motion of the breast itself, but rather it is produced by B0 field distortions associated with respiratory motion of tissues in the chest and abdomen. This susceptibility artifact has been reported to occur in the brain, but it is more apparent in the breast due to the anatomic proximity of the lungs. In the breast, these B0 distortions cause shot-to-shot frequency shifts, which vary an average of 24 Hz during a typical 1H MRS scan at 4 T. This variation can be corrected retrospectively by frequency shifting individual spectra prior to averaging. If not corrected, these shifts reduce spectral resolution and increase peak fitting errors. This work demonstrates the artifact, describes a method for correcting it, and evaluates its impact on quantitative spectroscopy. When the artifact is not corrected, quantification errors increase by an average of 28%, which dramatically impacts the ability to measure metabolite resonances at low signal-to-noise ratios. (c) 2004 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15562472     DOI: 10.1002/mrm.20277

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  18 in total

1.  Improved hyperthermia treatment control using SAR/temperature simulation and PRFS magnetic resonance thermal imaging.

Authors:  Zhen Li; Martin Vogel; Paolo F Maccarini; Vadim Stakhursky; Brian J Soher; Oana I Craciunescu; Shiva Das; Omar A Arabe; Williams T Joines; Paul R Stauffer
Journal:  Int J Hyperthermia       Date:  2010-11-11       Impact factor: 3.914

2.  Respiration artifact correction in three-dimensional proton resonance frequency MR thermometry using phase navigators.

Authors:  Bryant T Svedin; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2015-08-13       Impact factor: 4.668

3.  On restoring motion-induced signal loss in single-voxel magnetic resonance spectra.

Authors:  Refaat E Gabr; Shashank Sathyanarayana; Michael Schär; Robert G Weiss; Paul A Bottomley
Journal:  Magn Reson Med       Date:  2006-10       Impact factor: 4.668

Review 4.  Metabolite quantification and high-field MRS in breast cancer.

Authors:  Ihab S Haddadin; Adeka McIntosh; Sina Meisamy; Curt Corum; Angela L Styczynski Snyder; Nathaniel J Powell; Michael T Nelson; Douglas Yee; Michael Garwood; Patrick J Bolan
Journal:  NMR Biomed       Date:  2009-01       Impact factor: 4.044

5.  Correction of breathing-induced errors in magnetic resonance thermometry of hyperthermia using multiecho field fitting techniques.

Authors:  Cory R Wyatt; Brian J Soher; James R MacFall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

6.  MR spectroscopy of breast cancer for assessing early treatment response: Results from the ACRIN 6657 MRS trial.

Authors:  Patrick J Bolan; Eunhee Kim; Benjamin A Herman; Gillian M Newstead; Mark A Rosen; Mitchell D Schnall; Etta D Pisano; Paul T Weatherall; Elizabeth A Morris; Constance D Lehman; Michael Garwood; Michael T Nelson; Douglas Yee; Sandra M Polin; Laura J Esserman; Constantine A Gatsonis; Gregory J Metzger; David C Newitt; Savannah C Partridge; Nola M Hylton
Journal:  J Magn Reson Imaging       Date:  2016-12-16       Impact factor: 4.813

7.  Patient-to-patient variation of susceptibility-induced B₀ field in bilateral breast MRI.

Authors:  Seung-Kyun Lee; Ileana Hancu
Journal:  J Magn Reson Imaging       Date:  2012-06-11       Impact factor: 4.813

Review 8.  Biochemical characterization of breast tumors by in vivo and in vitro magnetic resonance spectroscopy (MRS).

Authors:  Uma Sharma; Naranamangalam R Jagannathan
Journal:  Biophys Rev       Date:  2009-01-17

9.  Accurate MR thermometry by hyperpolarized 129 Xe.

Authors:  Le Zhang; Alex Burant; Andrew McCallister; Victor Zhao; Karl M Koshlap; Simone Degan; Michael Antonacci; Rosa Tamara Branca
Journal:  Magn Reson Med       Date:  2016-10-19       Impact factor: 4.668

10.  Detection of choline signal in human breast lesions with chemical-shift imaging.

Authors:  Hyeon-Man Baek; Jeon-Hor Chen; Hon J Yu; Rita Mehta; Orhan Nalcioglu; Min-Ying Su
Journal:  J Magn Reson Imaging       Date:  2008-05       Impact factor: 4.813

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.