D I Thwaites1, J R Williams, E G Aird, S C Klevenhagen, P C Williams.
Abstract
A dosimetry intercomparison has been carried out for all 64 radiotherapy centres in the UK. Doses were measured with an ionization chamber in an epoxy resin water-substitute phantom of relatively simple geometry. Reference-point measurements were made for all MV photon beams. For 61 Co-60 beams, a mean ratio of measured-to-stated dose of 1.002 was observed with a standard deviation of 0.014, whilst for 100 MV x-ray beams, the corresponding figures were 1.003 and 0.015. 97% of beams lay within a +/- 3% deviation. One measurement was instrumental in discovering a large discrepancy. Doses were also investigated in two planned three-field distributions at one beam quality in each centre. One of these was in a homogeneous phantom, whilst the second included a lung-equivalent insert. Doses were measured at the central point and at four other points in the high dose volume. In both situations, the mean ratio of measured-to-calculated doses for all points was 1.008, with standard deviations of 0.027 and 0.035 for the uniform and non-uniform phantoms, respectively. Discrepancies over 5% were followed up. The work must be viewed in the context of other international intercomparisons and is an essential part of wider radiotherapy audit processes.Entities: