Literature DB >> 15519587

Thermal analysis of bone cement polymerisation at the cement-bone interface.

M Stańczyk1, B van Rietbergen.   

Abstract

The two major problems that have been reported with the use of polymethylmethacrylate (PMMA) cement are thermal necrosis of surrounding bone due to the high heat generation during polymerisation and chemical necrosis due to unreacted monomer release. Computer models have been used to study the temperature and monomer distribution after cementation. In most of these models, however, polymerisation is modelled as temperature independent and cancellous bone is modelled as a continuum. Such models thus cannot account for the expected important role of the trabecular bone micro-structure. The aim of this study is to investigate the distribution of temperature and monomer leftover at the cancellous bone-cement interface during polymerisation for a realistic trabecular bone-cement micro-structure and realistic temperature-dependent polymerisation kinetics behaviour. A 3-D computer model of a piece of bovine cancellous bone that underwent pressurization with bone-cement was generated using a micro-computed tomography scanner. This geometry was used as the basis for a finite element model and a temperature-dependent problem for bone cement polymerisation kinetics was solved to simulate the bone cement polymerisation process in the vicinity of the interface. The transient temperature field throughout the interface was calculated, along with the polymerisation fraction distribution in the cement domain. The calculations revealed that the tips of the bone trabeculae that are embedded in the cement attain temperatures much higher than the average temperature of the bone volume. A small fraction of the bone (10%) is exposed to temperatures exceeding 70 degrees C, but the exposure time to these high temperatures is limited to 50s. In the region near the bone, the cement polymerisation fraction (about 84%) is less than that in the centre (where it is reaching values of over 96%). An important finding of this study thus is the fact that the bone tissue that is subjected to the highest temperatures is also subjected to high leftover monomer concentration. Furthermore the maximum bone temperature is reached relatively early, when monomer content in the neighbouring cement is still quite high.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15519587     DOI: 10.1016/j.jbiomech.2004.03.002

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  27 in total

1.  Heat distribution of polymerisation temperature of bone cement on the spinal canal during vertebroplasty.

Authors:  Bernd Wegener; Nicole Zolyniak; Mehmet F Gülecyüz; Andreas Büttner; Christoph von Schulze Pellengahr; Veronika Schaffer; Volkmar Jansson; Christof Birkenmaier
Journal:  Int Orthop       Date:  2011-10-29       Impact factor: 3.075

Review 2.  Antibiotic-loaded Bone Cement as Prophylaxis in Total Joint Replacement.

Authors:  Javier Martínez-Moreno; Virginia Merino; Amparo Nácher; José Luis Rodrigo; Mónica Climente; Matilde Merino-Sanjuán
Journal:  Orthop Surg       Date:  2017-11       Impact factor: 2.071

3.  Dual setting α-tricalcium phosphate cements.

Authors:  T Christel; M Kuhlmann; E Vorndran; J Groll; U Gbureck
Journal:  J Mater Sci Mater Med       Date:  2012-12-13       Impact factor: 3.896

4.  HEAT TRANSFER MODEL AND QUANTITATIVE ANALYSIS OF DEEP TISSUE INJURY.

Authors:  Arjun Chanmugam; Akanksha Bhargava; Cila Herman
Journal:  Int Mech Eng Congress Expo       Date:  2012-11

5.  In vitro measurement of temperature changes during implantation of cemented glenoid components.

Authors:  Patric Raiss; Guido Pape; Sebastian Jäger; Markus Loew; Rudi Bitsch; Markus Rickert
Journal:  Acta Orthop       Date:  2010-04       Impact factor: 3.717

6.  Fluid-structure interactions in micro-interlocked regions of the cement-bone interface.

Authors:  Kenneth A Mann; Mark A Miller
Journal:  Comput Methods Biomech Biomed Engin       Date:  2013-03-12       Impact factor: 1.763

7.  Changes in microgaps, micromotion, and trabecular strain from interlocked cement-trabecular bone interfaces in total knee replacements with in vivo service.

Authors:  Mark A Miller; Jacklyn R Goodheart; Benjamin Khechen; Dennis Janssen; Kenneth A Mann
Journal:  J Orthop Res       Date:  2015-12-10       Impact factor: 3.494

8.  Recollapse of previous vertebral compression fracture after percutaneous vertebroplasty.

Authors:  D H Heo; D K Chin; Y S Yoon; S U Kuh
Journal:  Osteoporos Int       Date:  2008-07-18       Impact factor: 4.507

9.  Mechanical and thermal behaviour of an acrylic bone cement modified with a triblock copolymer.

Authors:  E Paz; J Abenojar; Y Ballesteros; F Forriol; N Dunne; J C Del Real
Journal:  J Mater Sci Mater Med       Date:  2016-02-17       Impact factor: 3.896

10.  Prevention of Oxygen Inhibition of PolyHIPE Radical Polymerization using a Thiol-based Crosslinker.

Authors:  Michael E Whitely; Jennifer L Robinson; Melissa C Stuebben; Hannah A Pearce; Madison A P McEnery; Elizabeth Cosgriff-Hernandez
Journal:  ACS Biomater Sci Eng       Date:  2017-01-23
View more