Evaluation of microfluidic blood gas sensors that combine microdialysis and optical monitoring.

C G Cooney1, B C Towe.   

Abstract

It is shown that microdialysis-based blood gas (pH, pCO2 and pO2) optical sensors are stable for durations of several hours in blood. This performance is uncommon with many types of membrane sensor. Microdialysis techniques can be designed to ensure that the sweep microflow samples are in biochemical equilibrium with the bulk media, even after hours of exposure to the complex composition of blood. The rate of diffusion through the membrane is not the determining factor in sensor reading, as it is with other sensor techniques that consume the analyte. The sweep fluid 95% equilibration times for microdialysis fibres were approximately double in blood compared with buffer, reflecting slower diffusion of ions. This is in contrast to the equilibration of gases through silicone hollow-fibre membranes in blood, which showed unchanged equilibration times between blood and buffer. Sensor measurements correlate well with a blood gas analyser for up to 9 h in blood, with correlation coefficients of 0.973 for the pO2 sensor 0.974 for the pCO2 sensor and 0.947 for the pH sensor. In blood, the sensors have precisions of 1.7 mmHg, 3.7 mmHg and 0.019 pH units and bias levels of -0.7 mmHg, 1.2 mmHg and 0.002 pH units, for pO2, pCO2 and pH, respectively.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15503975     DOI: 10.1007/bf02347556

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  9 in total

Review 1.  Applying the right statistics: analyses of measurement studies.

Authors:  J M Bland; D G Altman
Journal:  Ultrasound Obstet Gynecol       Date:  2003-07       Impact factor: 7.299

2.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

3.  Color flow Doppler: imaging the upper extremity.

Authors:  S Trager; M Pignataro; J Anderson; J M Kleinert
Journal:  J Hand Surg Am       Date:  1993-07       Impact factor: 2.230

4.  Bio-/haemocompatibility: implications and outcomes for sensors?

Authors:  M Kyrolainen; P Rigsby; S Eddy; P Vadgama
Journal:  Acta Anaesthesiol Scand Suppl       Date:  1995

5.  Effect of pH and buffering condition on dMPV of three types of platelet concentrates.

Authors:  M J Seghatchian; M E Alfonso; M V Vickers
Journal:  Transfus Sci       Date:  1997-03

6.  Intravascular carbon dioxide monitoring using micro-flow colorimetry.

Authors:  C G Cooney; B C Towe
Journal:  Biosens Bioelectron       Date:  1997       Impact factor: 10.618

Review 7.  The economic and clinical efficiency of point-of-care testing for critically ill patients: a decision-analysis model.

Authors:  M T Halpern; C S Palmer; K N Simpson; F D Chesley; B R Luce; J P Suyderhoud; B V Neibauer; F G Estafanous
Journal:  Am J Med Qual       Date:  1998       Impact factor: 1.852

8.  Value of point-of-care blood testing in emergent trauma management.

Authors:  A W Asimos; M A Gibbs; J A Marx; D G Jacobs; R J Erwin; H J Norton; M Thomason
Journal:  J Trauma       Date:  2000-06

9.  Prehospital point of care testing of blood gases and electrolytes - an evaluation of IRMA.

Authors: 
Journal:  Crit Care       Date:  1997       Impact factor: 9.097

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.