Centrically reordered inversion recovery half-Fourier single-shot turbo spin-echo sequence: improvement of the image quality of oxygen-enhanced MRI.

Yoshiharu Ohno1, Hiroto Hatabu, Takanori Higashino, Hideaki Kawamitsu, Hirokazu Watanabe, Daisuke Takenaka, Marc van Cauteren, Kazuro Sugimura.   

Abstract

PURPOSE: The purpose of the study presented here was to determine the improvement in image quality of oxygen-enhanced magnetic resonance (MR) subtraction imaging obtained with a centrically reordered inversion recovery half-Fourier single-shot turbo spin-echo (c-IR-HASTE) sequence compared with that obtained with a conventional sequentially reordered inversion recovery single-shot HASTE (s-IR-HASTE) sequence for pulmonary imaging.
MATERIALS AND METHODS: Oxygen-enhanced MR imaging using a 1.5 T whole body scanner was performed on 12 healthy, non-smoking volunteers. Oxygen-enhanced MR images were obtained with the coronal two-dimensional (2D) c-IR-HASTE sequence and 2D s-IR-HASTE sequence combined with respiratory triggering. For a 256x256 matrix, 132 phase-encoding steps were acquired including four steps for phase correction. Inter-echo spacing for each sequence was 4.0 ms. The effective echo time (TE) for c-IR-HASTE was 4.0 ms, and 16 ms for s-IR-HASTE. The inversion time (TI) was 900 ms. To determine the improvement in oxygen-enhanced MR subtraction imaging by c-IR-HASTE, CNRs of subtraction image, overall image quality, and image degradation of the c-IR-HASTE and s-IR-HASTE techniques were statistically compared.
RESULTS: CNR, overall image quality, and image degradation of c-IR-HASTE images showed significant improvement compared to those s-IR-HASTE images (P<0.05).
CONCLUSION: Centrically reordered inversion recovery half-Fourier single-shot turbo spin-echo (c-IR-HASTE) sequence enhanced the signal from the lung and improved the image quality of oxygen-enhanced MR subtraction imaging.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15489080     DOI: 10.1016/j.ejrad.2004.02.007

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  3 in total

1.  Oxygen-enhanced lung magnetic resonance imaging: influence of inversion pulse slice selectivity on inversion recovery half-Fourier single-shot turbo spin-echo signal.

Authors:  Daisuke Takenaka; Michael Puderbach; Yoshiharu Ohno; Frank Risse; Sebastian Ley; Kazuro Sugimura; Hans-Ulrich Kauczor
Journal:  Jpn J Radiol       Date:  2011-05-24       Impact factor: 2.374

2.  Vertical distribution of specific ventilation in normal supine humans measured by oxygen-enhanced proton MRI.

Authors:  Rui Carlos Sá; Matthew V Cronin; A Cortney Henderson; Sebastiaan Holverda; Rebecca J Theilmann; Tatsuya J Arai; David J Dubowitz; Susan R Hopkins; Richard B Buxton; G Kim Prisk
Journal:  J Appl Physiol (1985)       Date:  2010-10-07

3.  Oxygen-enhanced functional lung imaging using a contemporary 0.55 T MRI system.

Authors:  Ipshita Bhattacharya; Rajiv Ramasawmy; Ahsan Javed; Marcus Y Chen; Thomas Benkert; Waqas Majeed; Robert J Lederman; Joel Moss; Robert S Balaban; Adrienne E Campbell-Washburn
Journal:  NMR Biomed       Date:  2021-06-02       Impact factor: 4.478