New CR system with pixel size of 50 microm for digital mammography: physical imaging properties and detection of subtle microcalcifications.

Tadamitsu Ideguchi1, Yoshiharu Higashida, Yasuyuki Kawaji, Mitsuo Sasaki, Makoto Zaizen, Rei Shibayama, Yumiko Nakamura, Kimihiko Koyanagi, Hirotaka Ikeda, Masafumi Ohki, Fukai Toyofuku, Toru Muranaka.   

Abstract

PURPOSE: To investigate the physical imaging properties and detection of simulated microcalcifications of a new computed radiography (CR) system with a pixel size of 50 microm for digital mammography.
MATERIALS AND METHODS: New and conventional CR were employed in this study. The new CR system included a high-resolution imaging plate coupled with the FCR5000MA (50 microm pixel pitch) including transparent support and a dual-sided reader. The conventional CR system was coupled with the FCR9000 (100 microm pixel pitch). Modulation transfer functions (MTFs) and Wiener spectra (WS) of the new and conventional CR systems were measured. Observer performance tests were conducted to compare the effects of pixel size (50 microm vs. 100 microm) on the diagnostic accuracy of CR systems in the detection of simulated microcalcifications.
RESULTS: The presampling MTF of the new CR system was higher at high frequencies than the conventional CR system. The WS of the new CR system was comparable to that of the conventional CR system at all frequencies. The area under the receiver operating characteristic (ROC) curve (Az) obtained with the new CR and the conventional CR systems were 0.84 and 0.79, respectively. Results showed that the detection of simulated clustered microcalcifications was significantly improved by use of the new CR system compared with the conventional CR system (p<0.05).
CONCLUSION: The new CR mammography system improved physical imaging properties and detection of simulated microcalcifications over conventional CR mammography.

Entities:  

Mesh:

Year:  2004        PMID: 15468941

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  6 in total

1.  Diagnostic quality of 50 and 100 μm computed radiography compared with screen-film mammography in operative breast specimens.

Authors:  C M Pagliari; T Hoang; M Reddy; L S Wilkinson; J D Poloniecki; R M Given-Wilson
Journal:  Br J Radiol       Date:  2011-11-17       Impact factor: 3.039

Review 2.  [Clinical results of digital mammography].

Authors:  R Schulz-Wendtland; K-P Hermann; W Bautz
Journal:  Radiologe       Date:  2005-03       Impact factor: 0.635

Review 3.  Digital mammography: what do we and what don't we know?

Authors:  Ulrich Bick; Felix Diekmann
Journal:  Eur Radiol       Date:  2007-02-14       Impact factor: 5.315

4.  Digital magnification mammography with matched incident exposure: physical imaging properties and detectability of simulated microcalcifications.

Authors:  Nobukazu Tanaka; Kentaro Naka; Hiroko Fukushima; Junji Morishita; Fukai Toyofuku; Masafumi Ohki; Yoshiharu Higashida
Journal:  Radiol Phys Technol       Date:  2011-03-18

5.  Dose reduction in general radiography for adult patients by use of a dual-side-reading photostimulable phosphor plate in a computed radiography system.

Authors:  Norisato Tsuda; Nobukazu Tanaka; Tsutomu Akasaka; Hidetake Yabuuchi; Junji Morishita
Journal:  Radiol Phys Technol       Date:  2014-05-21

6.  Comparison of Image Quality Criteria between Digital Storage Phosphor Plate in Mammography and Full-Field Digital Mammography in the Detection of Breast Cancer.

Authors:  Pushpa Thevi Rajendran; Vijayalakshmi Krishnapillai; Sulaiman Tamanang; Kanaga Kumari Chelliah
Journal:  Malays J Med Sci       Date:  2012-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.