Literature DB >> 15465544

Comparison of posterior capsule opacification between the 1-piece and 3-piece Acrysof intraocular lenses: two-year results of a randomized trial.

Stefan Sacu1, Oliver Findl, Rupert Menapace, Wolf Buehl, Mathias Wirtitsch.   

Abstract

PURPOSE: To compare intensity of posterior capsule opacification (PCO) between the 1-piece and 3-piece haptic designs of an open-loop hydrophobic acrylic intraocular lens (IOL).
DESIGN: A randomized, patient- and examiner-masked clinical trial with intraindividual comparison. PARTICIPANTS: Fifty-two patients with bilateral age-related cataract (104 eyes).
METHODS: Each patient had cataract surgery in both eyes and received a 1-piece Acrysof IOL in one eye and a 3-piece Acrysof IOL in the fellow eye. Follow-up examinations were at 1 week, 1 month, 6 months, 1 year, and 2 years. Patients were examined at the slit lamp, visual acuity (VA) was determined, and standardized high-resolution digital retroillumination images of the posterior capsule were taken. The intensity of PCO was assessed subjectively at the slit lamp and objectively using automated image analysis software. MAIN OUTCOME MEASURE: Posterior capsule opacification score (scale, 0-10).
RESULTS: There was no significant difference between IOL styles in best-corrected VA, rhexis/IOL overlap, capsular folds, and amount of anterior capsule opacification during the follow-up period. One year postoperatively, the amount of regeneratory PCO was higher for the 1-piece Acrysof eyes (image analysis software score: 1.3) than for the 3-piece Acrysof eyes (score: 0.9; P = 0.002). However, 2 years postoperatively, there was no significant difference between the 2 IOL styles (1-piece: 1.5; 3-piece: 1.3; P = 0.3). Neodymium:yttrium-aluminum-garnet capsulotomy was not performed in the study.
CONCLUSION: One year postoperatively, the 1-piece Acrysof showed slightly more regeneratory PCO than the 3-piece Acrysof. However, 2 years postoperatively, the barrier effect of the 1-piece design was comparable to that of the 3-piece haptic design, with low PCO intensity.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15465544     DOI: 10.1016/j.ophtha.2004.03.028

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  11 in total

1.  [Influence of intraocular lens material and design on the development of posterior capsule opacification].

Authors:  O Nishi
Journal:  Ophthalmologe       Date:  2005-06       Impact factor: 1.059

2.  Risk factors for posterior capsule opacification.

Authors:  S D McLeod
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

3.  Nd:YAG capsulotomy rates after use of the AcrySof acrylic three piece and one piece intraocular lenses.

Authors:  S I Mian; K Fahim; A Marcovitch; H Gada; D C Musch; A Sugar
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

4.  Influence of three-piece and single-piece designs of two sharp-edge optic hydrophobic acrylic intraocular lenses on the prevention of posterior capsule opacification: a prospective, randomised, long-term clinical trial.

Authors:  Reda Zemaitiene; Vytautas Jasinskas; Gerd U Auffarth
Journal:  Br J Ophthalmol       Date:  2006-11-23       Impact factor: 4.638

Review 5.  Intraocular lens optic edge design for the prevention of posterior capsule opacification after cataract surgery.

Authors:  Sophie Maedel; Jennifer R Evans; Annette Harrer-Seely; Oliver Findl
Journal:  Cochrane Database Syst Rev       Date:  2021-08-16

6.  Intraocular lenses for the treatment of age-related cataracts: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2009-10-01

7.  Intraocular lens-edge design and material factors contributing to posterior-capsulotomy rates: comparing Hoya FY60aD, PY60aD, and AcrySof SN60WF.

Authors:  Peter J Morgan-Warren; Jm Alaric Smith
Journal:  Clin Ophthalmol       Date:  2013-08-22

8.  Comparison of Nd: YAG capsulotomy rate between 1-piece and 3-piece acrylic intraocular lenses: A STROBE-compliant article.

Authors:  Sun Young Kim; Young Kwon Chung; Hye Young Shin; Mee Yon Lee; Young-Chun Lee; Su-Young Kim
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

Review 9.  Biomaterial Influence on Intraocular Lens Performance: An Overview.

Authors:  Cari Pérez-Vives
Journal:  J Ophthalmol       Date:  2018-03-15       Impact factor: 1.909

10.  A meta-analysis of Nd:YAG capsulotomy rates for two hydrophobic intraocular lens materials.

Authors:  Mark Von Tress; James S Marotta; Stephen S Lane; Ramesh Sarangapani
Journal:  Clin Ophthalmol       Date:  2018-06-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.