A dynamic supraclavicular field-matching technique for head-and-neck cancer patients treated with IMRT.

Jun Duan1, Sui Shen, Sharon A Spencer, Raef S Ahmed, Richard A Popple, Sung-Joon Ye, Ivan A Brezovich.   

Abstract

PURPOSE: The conventional single-isocenter and half-beam (SIHB) technique for matching supraclavicular fields with head-and-neck (HN) intensity-modulated radiotherapy (IMRT) fields is subject to substantial dose inhomogeneities from imperfect accelerator jaw/MLC calibration. It also limits the isocenter location and restricts the useful field size for IMRT. We propose a dynamic field-matching technique to overcome these limitations. METHODS AND MATERIALS: The proposed dynamic field-matching technique makes use of wedge junctions for the abutment of supraclavicular and HN IMRT fields. The supraclavicular field was shaped with a multileaf collimator (MLC), which was orientated such that the leaves traveled along the superoinferior direction. The leaves that defined the superior field border moved continuously during treatment from 1.5 cm below to 1.5 cm above the conventional match line to generate a 3-cm-wide wedge-shaped junction. The HN IMRT fields were optimized by taking into account the dose contribution from the supraclavicular field to the junction area, which generates a complementary wedge to produce a smooth junction in the abutment region. This technique was evaluated on a polystyrene phantom and 10 HN cancer patients. Treatment plans were generated for the phantom and the 10 patients. Dose profiles across the abutment region were measured in the phantom on films. For patient plans, dose profiles that passed through the center of the neck lymph nodes were calculated using the proposed technique and the SIHB technique, and dose uniformity in the abutment region was compared. Field mismatches of +/- 1 mm and +/- 2 mm because of imperfect jaw/MLC calibration were simulated, and the resulting dose inhomogeneities were studied for the two techniques with film measurements and patient plans. Three-dimensional volumetric doses were analyzed, and equivalent uniform doses (EUD) were computed. The effect of field mismatches on EUD was compared for the two match techniques.
RESULTS: For a perfect jaw/MLC calibration, dose profiles for the 10 patients in the 3-cm match zone had an average inhomogeneity range of -1.6% to +1.6% using the dynamic-matching technique and -3.7% to +3.8% according to the SIHB technique. Measurements showed that dose inhomogeneities that resulted from 1-mm and 2-mm jaw/MLC calibration errors were reduced from as large as 27% and 45% with the SIHB technique to less than 2% and 5.7% with the dynamic technique, respectively. For -1-mm, -2-mm, +1-mm, and +2-mm jaw/MLC calibration errors, respectively, treatment plans for the 10 patients yielded average dose inhomogeneities of -5.9%, -3.0%, +2.7%, and +5.8% with the dynamic technique as compared to -22.8%, -11.1%, +9.8%, and +22.1% with the SIHB technique. Calculation based on a dose-volume histogram (DVH) showed that the SIHB technique resulted in larger changes in EUD of the PTV in the junction area than did the dynamic technique.
CONCLUSION: Compared with the conventional SIHB technique, the dynamic field-matching technique provides superior dose homogeneity in the abutment region between the supraclavicular and HN IMRT fields. The dynamic feathering mechanism substantially reduces dose inhomogeneities that result from imperfect jaw/MLC calibration. In addition, isocenter location in the dynamic field-matching technique can be chosen for reproducible patient setup and for adequate IMRT field size rather than being dictated by the match position. It also allows angling of the supraclavicular field to reduce the volume of healthy lung irradiated, which is impractical with the SIHB technique. In principle, this technique should be applicable to any treatment site that requires the abutment of static and intensity-modulated fields.

Entities:  

Mesh:

Year:  2004        PMID: 15465215     DOI: 10.1016/j.ijrobp.2004.06.213

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  9 in total

1.  A novel dynamic field-matching technique for treatment of patients with para-aortic node-positive cervical cancer: Clinical experience.

Authors:  Craig Baden; Alexander Whitley; Javier López-Araujo; Richard Popple; Jun Duan; Robert Kim
Journal:  Rep Pract Oncol Radiother       Date:  2015-11-21

2.  Use of a conventional low neck field (LNF) and intensity-modulated radiotherapy (IMRT): no clinical detriment of IMRT to an anterior LNF during the treatment of head-and neck-cancer.

Authors:  Aruna Turaka; Tianyu Li; Nicos Nicolaou; Miriam N Lango; Barbara Burtness; Eric M Horwitz; John A Ridge; Steven J Feigenberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-10       Impact factor: 7.038

3.  Effects of full-neck volumetric-modulated arc therapy vs split-field intensity-modulated head and neck radiation therapy on low neck targets and structures.

Authors:  Shibu J Anamalayil; Boon-Keng K Teo; Alexander Lin; Robert A Lustig; Peter H Ahn
Journal:  Br J Radiol       Date:  2016-04-04       Impact factor: 3.039

4.  Verification of field match lines in whole breast radiation therapy using Cherenkov imaging.

Authors:  Rachael Hachadorian; J Cedar Farwell; Petr Bruza; Michael Jermyn; David J Gladstone; Brian W Pogue; Lesley A Jarvis
Journal:  Radiother Oncol       Date:  2021-05-01       Impact factor: 6.901

5.  Lung cancer: a 6-field technique using lateral beams in conformal radiotherapy for bilateral supraclavicular lymph node metastases.

Authors:  Shinichi Tsutsumi; Takuhito Tada; Tomoko Maekado; Masahiro Tokunaga; Noriko Tanaka; Ai Kobayashi; Eiichiro Okazaki; Shougo Matsuda; Masako N Hosono; Yukio Miki
Journal:  Springerplus       Date:  2014-12-13

6.  Total body irradiation with step translation and dynamic field matching.

Authors:  Ho-Hsing Chen; Jay Wu; Keh-Shih Chuang; Jia-Fu Lin; Jia-Cheng Lee; Jin-Ching Lin
Journal:  Biomed Res Int       Date:  2013-07-01       Impact factor: 3.411

7.  An opposed matched field IMRT technique for prostate cancer patients with bilateral prosthetic hips.

Authors:  Shahin Fattahi; Orest Z Ostapiak
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

8.  MLC-based penumbra softener of EDW borders to reduce junction inhomogeneities.

Authors:  Stanislaw Szpala; Kirpal Kohli
Journal:  J Appl Clin Med Phys       Date:  2017-04-19       Impact factor: 2.102

9.  Effect of interfractional shoulder motion on low neck nodal targets for patients treated using volumetric-modulated arc therapy (VMAT).

Authors:  Kevin E Casey; Pei-Fong Wong; Samuel S Tung
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.