Abstract

PURPOSE: To increase gradient stimulation from 100% to a fixed 120% level and to assess patient acceptance of the degree of peripheral nerve stimulation (PNS) at magnetic resonance (MR) imaging.
MATERIALS AND METHODS: Two hundred ten patients underwent MR imaging of various body regions according to clinical indications. An additional three-dimensional fast low-angle shot sequence with the 120% stimulation level was performed. A patient questionnaire was distributed after MR imaging to document the presence, degree, and location of PNS. Degree was measured with an 11-point scale (score range, 0-10). Age was analyzed between the sexes for significant statistical differences. Furthermore, correlation between location of examination and location and degree of stimulation was performed. To determine stimulation discomfort relative to other factors typically present at MR imaging, the degree of discomfort due to room temperature, size of magnet bore, acoustic noise, examination time, and heating sensation was determined for comparison, as well.
RESULTS: Thirty-five (16.7%) patients reported a stimulation sensation during imaging in one or more locations, while six (2.9%) felt very uncomfortable local stimulation (score of 8-10). No significant difference between male and female patients regarding age, sex, and appearance or degree of stimulation sensation could be detected. No significant correlation between location of examination and location and degree of stimulation was recorded. Compared with other side effects, PNS was considered relatively unimportant.
CONCLUSION: The 120% gradient stimulation level seems acceptable for routine clinical imaging with this gradient system, since only 2.9% of patients experienced very uncomfortable local stimulation.

Entities:  

Mesh:

Year:  2004        PMID: 15358851     DOI: 10.1148/radiol.2332030428

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  9 in total

Review 1.  [Physical interactions in MRI: Some rules of thumb for their reduction].

Authors:  M Mühlenweg; G Schaefers; S Trattnig
Journal:  Radiologe       Date:  2015-08       Impact factor: 0.635

2.  Subjective acceptance of 7 Tesla MRI for human imaging.

Authors:  Jens M Theysohn; Stefan Maderwald; Oliver Kraff; Christoph Moenninghoff; Mark E Ladd; Susanne C Ladd
Journal:  MAGMA       Date:  2007-12-07       Impact factor: 2.310

Review 3.  [Safety aspects in high-field magnetic resonance imaging].

Authors:  M Mühlenweg; G Schaefers; S Trattnig
Journal:  Radiologe       Date:  2008-03       Impact factor: 0.635

Review 4.  Risks and safety aspects related to PET/MR examinations.

Authors:  Gunnar Brix; Elke A Nekolla; Dietmar Nosske; Jürgen Griebel
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

Review 5.  Biological effects and safety in magnetic resonance imaging: a review.

Authors:  Valentina Hartwig; Giulio Giovannetti; Nicola Vanello; Massimo Lombardi; Luigi Landini; Silvana Simi
Journal:  Int J Environ Res Public Health       Date:  2009-06-10       Impact factor: 3.390

Review 6.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

7.  Induced current magnetic resonance electrical impedance tomography of brain tissues based on the J-substitution algorithm: a simulation study.

Authors:  Yang Liu; Shanan Zhu; Bin He
Journal:  Phys Med Biol       Date:  2009-06-26       Impact factor: 3.609

Review 8.  Occupational exposure to electromagnetic fields in magnetic resonance environment: an update on regulation, exposure assessment techniques, health risk evaluation, and surveillance.

Authors:  Valentina Hartwig; Giorgio Virgili; F Ederica Mattei; Cristiano Biagini; Stefania Romeo; Olga Zeni; Maria Rosaria Scarfì; Rita Massa; Francesco Campanella; Luigi Landini; Fabriziomaria Gobba; Alberto Modenese; Giulio Giovannetti
Journal:  Med Biol Eng Comput       Date:  2021-09-29       Impact factor: 2.602

9.  Vestibular effects of a 7 Tesla MRI examination compared to 1.5 T and 0 T in healthy volunteers.

Authors:  Jens M Theysohn; Oliver Kraff; Kristina Eilers; Dorian Andrade; Marcus Gerwig; Dagmar Timmann; Franz Schmitt; Mark E Ladd; Susanne C Ladd; Andreas K Bitz
Journal:  PLoS One       Date:  2014-03-21       Impact factor: 3.240