Literature DB >> 15348739

Comparison of the wear of aged and non-aged ultrahigh molecular weight polyethylene sterilized by gamma irradiation and by gas plasma.

J Fisher1, E A Reeves, G H Isaac, K A Saum, W M Sanford.   

Abstract

The wear of ultrahigh molecular weight polyethylene in artificial joints is a major cause of long-term osteolysis and loosening. The wear rate of aged and non-aged ultrahigh molecular weight polyethylene sterilized by both gamma irradiation in air and gas plasma has been studied in simple configuration wear tests. Fourier transform infrared analysis (FTIR) showed marked oxidative degradation of the irradiated and aged material compared to the gas plasma sterilized and aged material. The wear rate of the irradiated and aged material was significantly (three times) higher than the gas plasma sterilized, gas plasma sterilized and aged, and non-sterilized materials. Alternative sterilization procedures such as gas plasma, when used instead of gamma irradiation in air, are likely to reduce the incidence of long-term osteolysis.

Entities:  

Year:  1997        PMID: 15348739     DOI: 10.1023/a:1018537001366

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  3 in total

1.  Reactions of the articular capsule to wear products of artificial joint prostheses.

Authors:  H G Willert; M Semlitsch
Journal:  J Biomed Mater Res       Date:  1977-03

2.  Preliminary study of the effect of aging following irradiation on the wear of ultrahigh-molecular-weight polyethylene.

Authors:  J Fisher; K L Chan; J L Hailey; D Shaw; M Stone
Journal:  J Arthroplasty       Date:  1995-10       Impact factor: 4.757

3.  Property changes of UHMW polyethylene hip cup endoprostheses during implantation.

Authors:  P Eyerer; Y C Ke
Journal:  J Biomed Mater Res       Date:  1984 Nov-Dec
  3 in total
  4 in total

1.  The influence of gamma irradiation and aging on degradation mechanisms of ultra-high molecular weight polyethylene.

Authors:  F J Buchanan; J R White; B Sim; S Downes
Journal:  J Mater Sci Mater Med       Date:  2001-01       Impact factor: 3.896

2.  The Ringloc liner compared with the Hexloc liner in total hip arthroplasty.

Authors:  Sköldenberg Olof; Salemyr Mats; Muren Olle; Johansson Ake; Ahl Torbjörn; Bodén Henrik
Journal:  Orthop Rev (Pavia)       Date:  2009-10-10

3.  Influence of Accelerated Aging on the Wear Behavior of Cross-Linked Polyethylene Liners-A Hip Simulator Study.

Authors:  Rebecca H Dammer; Carmen Zietz; Jörn Reinders; Michael Teske; Rainer Bader
Journal:  J Clin Med       Date:  2022-08-01       Impact factor: 4.964

4.  (iv) Enhancing the safety and reliability of joint replacement implants.

Authors:  Louise M Jennings; Mazen Al-Hajjar; Claire L Brockett; Sophie Williams; Joanne L Tipper; Eileen Ingham; John Fisher
Journal:  Orthop Trauma       Date:  2012-08
  4 in total